59
59
59
SD GlucoNavii® NFC BLOOD GLUCOSE MONITORING SYSTEM
Main
tenanc
e and
T
roubleshooting
59
[Lo message]
t
Blood glucose may be lower than
the measuring range of the system.
See chapter 3 “4. Understanding
Test Results”.
[Internal Error Message for a
meter]
t
Turn off a meter. Then turn on
the meter again. If there is still
error message, please contact SD
Biosensor, Inc. TEL: +82-31-300-0400.
[Strip Error]
t
Defective test strip or the test strip
is damaged or inserted improperly.
Discard this test strip and test again
using new test strip. See Chapter 3 “2.
Performing a Blood Glucose Test”.