SCS 770030 User Manual Download Page 6

TB-9071 

Page 6  of 

© 2016 DESCO INDUSTRIES, INC. 

Employee Owned 

 

Calibration 

The SCS Combo Tester is calibrated to NIST traceable 

standards. There are no user adjustments necessary. 

We recommend that calibration be performed 

annually using the SCS 

770033 

Calibration Unit. It 

is a convenient product which allows the customer 

to perform NIST traceable calibration on single-wire 

personnel testers. The 

770033 

is designed to be used 

on the shop floor at the workstation, virtually eliminating 

downtime, verifying that the tester is operating within 

tolerances. 

View technical bulletin 

TB-9073 

for more information on 

the 

770033 

Calibration Unit and instructions to calibrate 

the Combo Tester. 

Specifications 

COMBO TESTER 

Power Supply 

9V Alkaline Battery or 

optional AC power adapter 

Operating Temperature 

32°F to 104°F (0 to 40°C) 

Wrist Strap Test Limit: 

750 kilohms to 10 megohms 

Footwear Test Limit: 

750 kilohms to 100 megohms 

Test Accuracy: 

±20% 

Test Voltage 

4 VDC 

Dimensions 

3.3" x 5.6" x 1.5" 

Weight 

0.5 lbs 

 

STAND AND FOOT PLATE 

Stand Height 

36" 

 

Foot Plate Dimensions 

18" x 12" x 1" 

Weight 

8.0 lbs 

 

STAINLESS STEEL FOOT PLATE 

Dimensions 

20" x 20" x 0.5" 

Weight 

5.5 lbs 

 
 
 
 
 
 

 

Figure 13. SCS 

770033 

Calibration Unit 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Texas Technologies, Inc. 

3600 W. Whitestone Blvd. Cedar Park, TX 78613

 

512.267.0100 

 

https://www.texastechnologies.com/static-control-2/testers-meters/ 