Maintenance 

Before disassembling make sure all fluid is drained from the meter and connecting pipes. Cleaning 
can be carried out using a soft brush. Be careful not to damage the chamber or disk. Use only 
original spare parts. A new calibration should be performed after cleaning or disassembly.  
 

Flow Meter Disassembly 

1) 

Remove the reset knob by pulling firmly to the side. 

2) 

Loosen the 4 retaining screws see Diagram 1, pos 22. 

3) 

Loosen the 2 screws at pos 5. 

4) 

Reassembly is the reverse. 

 

Measuring Chamber 

1) 

Disassemble the unit as per the above instructions. 

2) 

Loosen the 8 screws at pos 7. 

3) 

Remove the body cover, pos 28, together with the gear unit. Be careful not to damage the 

gasket, pos 16. 
4) 

Remove the whole measuring chamber, pos 27, by lifting it from the meter body and at the 

same time pulling it back towards the inlet in order to remove the O ring, pos 17, from its seal at 
the outlet. 
5) 

To check the inside of the measuring chamber, pos 27, remove the O ring, pos 17, and 

divide the two half chambers containing the nutating disks. 
6) 

To reassemble the chamber reverse the procedure. Be careful to verify the disk rotates 

freely in the chamber. Install the gaskets properly and make sure to lubricate them. Make sure the 
while fixing the cover on the body the nutating disk needle does not hit the gear, pos 14, which 
must remain free to be pulled by the disk needle. Tighten screws at pos 7. 
 

Gear Unit 

To reach the gear unit. 
1) 

Remove cover, pos 28. 

2) 

Loosen screws, pos 13 

3) 

Remove plate, pos 12. Now all gears can be reached for inspection. 

4) 

Should the gasket, pos 10, need to be replaced, remove the bevel gear, pos 8, from the 

shaft by pulling axially, then remove the gear, pos 11, together with the shaft. 
5) 

Reassembly is the reverse procedure. 

Summary of Contents for SPEP12V75

Page 1: ...st No Part Name QTY No Part Name QTY 1 Screw M8 16 2 14 Ring 1 2 Nozzle Holder 1 15 Cover 1 3 Motor 1 16 Screw M6 20 8 4 Pump Body 1 17 Inlet Joint 1 5 Bypass Valve 1 18 Filter 1 6 Spring 1 19 Gasket 1 7 O ring 26 5 2 65 1 20 Oil Pipe 8 Cover 1 21 Screw M5 12 4 9 Seals Assy 1 22 Cover 1 10 End Ring 9 1 23 O ring 1 11 Key 1 24 Upper LeverⅠ 1 12 Rotor 1 25 Upper Lever 2 1 ...

Page 2: ...Exploded Pump View Features Voltage 12V Power 185 Watts Duty Cycle 30 Minutes Flow Rate 20 75LPM Operating Pressure 0 14MPA Current Draw 15 A Inlet Outlet 1 BSP Drum Fitting 2 BSP Hose length 4 meters ...

Page 3: ... turned On or OFF with the Upper and Lower levers Move the Upper Lever to ON 6 Keep the nozzle in contact with the drum being filled Warning 1 The drum or container being filled should be earthed to prevent static discharge and possible fire or explosion The discharge hose is not conductive and does not provide an earth 2 Do not install additional foot valves or check valves that do not have a pre...

Page 4: ...uding temperature pressure and viscosity If the accuracy of the flow meter is compromised then a calibration should be performed Calibrating flow meters is a normal part of their maintenance and is not an indication of a fault with the unit Calibration procedure 1 Unscrew the brass calibration plug shown in diagram 1 2 Purge the system pump piping meter of air 3 Stop the flow by shutting off the n...

Page 5: ...dy and at the same time pulling it back towards the inlet in order to remove the O ring pos 17 from its seal at the outlet 5 To check the inside of the measuring chamber pos 27 remove the O ring pos 17 and divide the two half chambers containing the nutating disks 6 To reassemble the chamber reverse the procedure Be careful to verify the disk rotates freely in the chamber Install the gaskets prope...

Page 6: ......

Reviews: