PA: Polyallomer

PC

Polycarbonate

PP

Polypropylene 

11.2.2 Cleaning PC tubes

 

PC materials are low in chemical resistance against alkaline solutions. Avoid using neutral detergents with 

pH higher than 9. Note that pH of some neutral detergents are still higher than 9 even if diluted according to 

the instruction in the maker‟s catalog. Use detergent with its pH between 7 and 9. 

11.2.3 Autoclaving PA

PC and PP tubes

 

PA begins softening at about 120

, PC and PP at about 130

. Autoclave PA tubes at 115

℃(

0.7kg/cm

2

for

30 minutes and PC and PP tubes at 121

℃(

0.1kg/cm

2

for 20 minutes. If a certain temperature is exceeded,

the tubes may be deformed. 

When using a sterilizing chamber, please operate as follows: 

1

) 

Place tubes in vertical position, mouths upward. If tubes are placed sideways, they may deform

into an oval shape due to gravity.

2

) 

Remove screw nuts and inner covers to prevent from deformation or rupture.

3

) 

Wait  until  the  sterilizing  chamber  cools  down  to  the  room  temperature  before  the  tubes  are

removed.

11.2.4 Condition and life expectancy of tubes 

The life expectancy of plastic tubes depends on the characteristics of samples, speed of the rotor used, and 

temperature applied, and so on. When the plastic tubes are used for centrifuge of ordinary aqueous samples 

(pH between 5 and 9), their life expectancies are defined as follows. 

Be operated at the maximum speed: 

High quality tubes (PA

PC

PP): 30-50 operations   

Ordinary tubes(PA

PC

PP): around 10 operations

Using in low speed can extend the tube life

Life  expectancy  of  tubes  also  depends  on  the  pretreatment  conditions  such  as  cleaning  and  sterilization, 

lifetime can be cut down. 

Notice: Do not use damaged or cracked tubes. 

12. Calculate Relative Centrifuge Force(RCF)

Relative Centrifuge Force (RCF) can be determined with the following calculation formula. 

RCF=1.118

×

r

×

n

2

×

10

-5

R—rotating radius, unit: cm; n—rotating speed, unit: rpm 

Summary of Contents for SCI636

Page 1: ...SCI636 Multi purpose Centrifuge Before using centrifuge please carefully read this user manual for its efficient operation and safety...

Page 2: ...ect the rotor 8 7 5 Symmetrically load centrifuge tubes in rotor 8 8 Operation 8 8 1 Normal Operation 8 8 2 RCF Operation 12 8 3 Pulse Operation 12 9 Maintenance 12 9 1 Cleaning 12 9 2 Consumables 13...

Page 3: ...USER MANUAL 2...

Page 4: ......

Page 5: ...y be reproduced or transmitted without prior written permission of original manufacture We can not be responsible to inform at real time if the outline and specifications of centrifuge is subject to c...

Page 6: ...death CAUTION Possible damage to instrument Caution notes indicate any condition or practice which if not strictly observed or remedied could result in damage or destruction of the instrument NOTE Not...

Page 7: ...orizontal before running Make sure the angle between the door and cover is greater than 70 degrees when open the door Be careful not put your fingers or hands between the door and cover when the door...

Page 8: ...emperature detector Imbalance protection Automatic internal diagnosis Power requirements Single phase 220V 240V 50Hz 60Hz 8A 110V 120V 50Hz 60Hz 12A Dimensions mm L 280 W 364 H 266 Weight 30kg Additio...

Page 9: ...ich case the user will be required to correct the interference 3 Required Operational Condition 3 1 Basic operational Conditions 1 Power Single phase 220V 240V 50Hz 60Hz 8A Single phase 110V 120V 50Hz...

Page 10: ...at least 30cm behind it to guarantee the cooling efficiency 4 Keep away from heat or water to avoid sample temperature issues or centrifuge failures 4 2 Connection of the power cord and grounding WAR...

Page 11: ...Open lock key Press the key to open the door The key is not available when the centrifuge is running 3 Start Stop key Press the key to start running The centrifuge will brake to stop running if press...

Page 12: ...is 9 memory program is 7 When speed symbol is rotating indicating the centrifuge is running the rotation is faster the speed is higher Temperature only displays the temperature of chamber and can not...

Page 13: ...erwise the rotor may be moved off while rotating and cause damage of the centrifuge and rotor Firmly tighten the rotor door with rotor 8 Operation CAUTION Do not push or lean against the centrifuge wh...

Page 14: ...00rpm Running time 30 minutes The door lock is released 8 1 1 Load and replace the rotor Figure 8 2 Load the rotor CAUTION Attach the rotor to the rotor shaft Ensure the rotor is in position and conne...

Page 15: ...peed symbol will flash the speed value The minimum speed value you can set 300rpm the minimum increment is 100rpm Rotate parameter key clockwise to increase speed value rotate the parameter key anti c...

Page 16: ...normal operation mode and run according to the new value If the set time value has been modified the operation time is not affected and will continue 3 Warning display If an error occurs during the o...

Page 17: ...the power switch and load the rotor to the shaft tighten the rotor lid and make sure it is in secured position and then close the door 2 The centrifuge gets into preparation mode and displays last ru...

Page 18: ...mple leaks in the rotor rinse the rotor with water Apply a thin coat of silicon grease to the rotor when it is completely dry The rotor should be regular maintenance recommend to cleaning it each 3 mo...

Page 19: ...n the table firmly 2 Check if the centrifuge grounded properly Use multi meter to check if is short circuit between the power cord grounding pin and the motor shaft If yes indicating grounded properly...

Page 20: ...llowable imbalance is exceeded some wrong in the drive system Balance the sample checked by professional person E 10 86 Read the service manual Contact with service center Table 10 1 Possible problems...

Page 21: ...eed the allowable maximum speed of rotor tube and adapters etc be care that the allowable maximum speed of some adapters are lower than the rotor s maximum speed 11 1 The rotor instructions 11 1 1 Rot...

Page 22: ...50mlV 4 A50VP100 4000 15 9 2840 15mlV 8 A15VP100 4000 15 1 2700 3 10ml 8 A10P100 4000 14 7 2630 9 S2 MP dimension mm MTP 128 85 6 15 4000 12 1 2160 Cell culture 128 85 6 21 4000 12 1 2160 DWP 128 85 6...

Page 23: ...5 Bio safe seal ring The rotor is sealed by bio safe structures achieved using three high temperature rubber seal rings The seal rings may fall off or aging after several autoclaving need to be repla...

Page 24: ...ove screw nuts and inner covers to prevent from deformation or rupture 3 Wait until the sterilizing chamber cools down to the room temperature before the tubes are removed 11 2 4 Condition and life ex...

Page 25: ...Rotor kits with cover fixing clips and O ring 6000rpm PCR8 12 used with D1536 D1536R 19400002 AS6 50V Rotor kits with cover fixing clips and O ring 6000rpm 50mlV 6 used with D1536 D1536R 19400002 A30...

Page 26: ...the following conditions even if within the guarantee period expires 1 Failures caused by incorrect installation 2 Failures caused by rough or improper handling 3 Failures caused by conveyance or rel...

Page 27: ...USER MANUAL 22 SCILOGEX LLC 1275 Cromwell Ave Suite C6 Rocky Hill CT 06067 United States Office 1 860 436 9221 Fax 1 860 436 9745 Sales service contact info scilogex com www scilogex com...

Reviews: