USER MANUAL
(4) Set program
There are 1
~
9 program groups.
Saving the program
Press and hold the key
more than 5 seconds, the current parameters are saved under the selected
program number.
Loading the program
Press the key
shortly, the program number will be increasing, from 1 to 9, then from 9 to 1, the
corresponding parameters changing as well.
8.1.4 Start the operation
(
1
)
Press key
to start running
The door should be locked before rotor starts rotating.
Timer will operate once the speed setting value is reached, the screen displays the remaining run time.
(
2
)
View and modify the operation programs
Pressing key
, returns the display to the program interface and displays settings programs. Press the
key
to the desired program. When flashing, rotate the key
to modify values. Release the key
after 5 seconds, and the centrifuge will return to normal operation mode and run according to the new
value.
If the set time value has been modified, the operation time is not affected and will continue.
(
3
)
Warning display
If an error occurs during the operation, the centrifuge will brake to stop automatically, and display the
error code on the time/display area. The error code can be checked in the table 10-1, and corrective
actions can be applied accordingly.
8.1.5 End the operation
(
1
)
The centrifuge will brake when it reaches the setting time or
key is pressed.
When the rotor stops rotating, centrifuge will start beeping to alert the operation has finished.
(
2
)
Open the door
The door can be released automatically when the operation has stopped (D1536).
The centrifuge keeps the door close when operation has stopped (D1536R).
With the door closed, you are able to press the
key to open it.
After ending the operation, the program will store the setting parameters of this operation, and will
recall these parameters when restart the program.
(
3
)
Open the door and take out the rotor and samples.