USER MANUAL
5
Deceleration key
Press the key to set the deceleration curve.
0: inertial braking. 9: fastest braking.
6
Memory key
Load program: Press the key shortly to load the program
Save program: Press and hold key for 5 seconds
7
Parameter key
Clockwise rotate to increase parameter values. counter-
clockwise rotate to decrease parameter values.
Press down the key, shift between speed , RCF and time
display.
Figure 6-2 the main interface
Main interface is as figure 6-2. The speed is set to be 6000rpm, temperature of
centrifugal chamber is 20
℃
,
and the running time is 30 minutes, acceleration curve is 9, and deceleration curve is 9, memory program is
7. When speed symbol
is rotating, indicating the centrifuge is running, the rotation is faster, the speed is
higher. Temperature only displays the temperature of chamber and can not be controlled. Time symbol
displays the ratio of working to time setting. The total time setting is divided into 10 scales.
7. Rotor Preparation
7.1 Prepare the samples
7.2 Inject the samples into tubes.
CAUTION
:
Do not overload samples into the centrifuge which will cause leaking.
Do not exceed the actual capacity allowed in the user manual.
7.3 Keep the tubes balance
Although the centrifuge can accept sample balancing by eye, we recommend that you keep this
centrifuge in a well-balanced condition to extend its life expectancy.
Never intentionally run the centrifuge under unbalanced condition even though the allowable