Rotary Evaporator
•
Press Lift-up/Lift-down key, the lift goes up/down.
Release Lift-up/Lift-down key, the lift stops.
•
Push the speed key and then rotate control knob to set
the rated speed.
•
Push the temperature key and then rotate control knob
to set the rated temperature.
•
Push the timer key and then rotate control knob to set
the rated time.
Deactivate the timer by setting the target value to “0”.
•
Press the Int key and then rotate control knob to set the
rated interval value.
Deactivate the interval by setting the target value to
“0”.
•
Press the control knob, switch ON.
If these operations above are normal, the device is ready to
operate. If not, the device may be damaged during
transportation, please contact manufacturer/supplier for
technical support.
Warning !
Forbid to transfer the flask when the instrument
working.
6.
Faults
•
Instruments can’t be power ON
-
Check whether the power line is unplugged
-
Check whether the fuse is broken or loose
•
Fault in power on self test
-
Switch OFF the unit, then switch ON and reset
the instruments to factory default setting.
•
ERR 003, lift operation timeout
-
Load overweight
•
ERR 005, lift can’t work
-
Check whether the transportation lock is loosen
-
Load overweight
If these faults are not resolved, please contact
manufacturer/supplier.