![Scilogex SCI100-Pro User Manual Download Page 13](http://html1.mh-extra.com/html/scilogex/sci100-pro/sci100-pro_user-manual_1217654013.webp)
11
Rotary Evaporator
Figure 8
•
Open the locking device on the drive head by turning
it 60° counter clockwise (e), (Figure 9).
•
Feed the steam pipe in until it stops.
•
Then lock the locking device by turning it clockwise
by 60°.
•
The plastic screw nut (f) helps loosen tight-fitting
piston ground-in connections (Figure 9).
-
Hold the tight-fitting evaporator piston and
counter clockwise (e), and then turn the plastic
screw nut (f) until the evaporator piston neck.
Figure 9
Note: Prior to commissioning, hand-tighten the plastic
screw nut (f) left-aligned.
•
Installing condenser seal (Figure 10 and Figure 11).
Figure 10
d
f
e