6.

 

 Function: Heating    

The device is controlled by digital temperature control 

technology, and it has a over temperature protection, the 

safety temperature is 140

 

Press the temperature switch button, the characters in 

temperature display area will be flashing, and then rotate 

the parameter setting knob to set the target temperature, 

press the parameter setting knob to turn on the setting. 

Note: 

 

Set the temperature via rotating the parameter setting 

knob slowly to the target value. 

 

When the heating function is switched on, the LED 
displays the temperature value on the left-hand side.   

 

The heating function is switched on or off by pushing 

heating knob.

 

The instrument automatically displays the last running 

speed and temperature parameters once turned on.   

Generally, the LED screen cannot display the actual 

temperature of sample in the vessel or dry bath surface,   

temperature differences as following: 

 

Dry bath center and outer edge. 

 

The sample in the vessel and dry bath surface. 

7.

 

 Function: Timing 

The device can work in accordance with the timing mode or 

continuous operation mode. 

Timing Mode: Press the time switch button, the characters 

in time display area will be flashing, and then rotate the 

parameter setting knob to set the target time, press the 

parameter setting knob to turn on the setting. 

During operation, motion can be stopped at any time by 

pressing the speed/timer control knob. If the knob is 

pressed again, motion will start again and the timer will

 

restart countdown. When the timer reaches zero, the unit 

will be automatically halted. 

Continuous Mode: Press the time switch button, the 

characters in time display area will be flashing, make sure 

the setting time is zero, press the parameter setting knob to 

turn on the continuous operation mode. 

 

 

 

 

 

 

Summary of Contents for HB120-S

Page 1: ...B120 S LED Digital Dry Bath Please read the User Manual carefully before use and follow all operating and safety instructions Technical specifications and outline are subject to change without prior n...

Page 2: ...anty 1 Safety Instructions 2 Proper Use 3 Inspection 4 Control 5 Trial Run 6 Function Heating 7 Function timing 8 Faults 9 Maintenance and Cleaning 10 Associated Standards and Regulations 11 Specifica...

Page 3: ...em i e hardware or software Methods and procedures adopted to resolve the problems Your contact information Warranty Changes or modifications not expressly approved by the party responsible for compli...

Page 4: ...nths from the date of invoice The warranty is extended only to the original purchaser It shall not apply to any product or parts which have been damaged on account of improper installation improper co...

Page 5: ...es or under water Temperature must always be set at least 50 lower than the fire point of the media used Be aware of hazards due to Flammable materials or media with a low boiling temperature Overfill...

Page 6: ...n residential areas or other constraints mentioned in Chapter 1 3 Inspection 3 1 Receiving Inspection Unpack the equipment carefully and check for any damages which may have arisen during transport Pl...

Page 7: ...ch button Press the button can set the target temperature Time switch button Press the button can set the target time LED Display LED displays the real working state and all settings LED Lignt The LED...

Page 8: ...in time display area will be flashing and then rotate the parameter setting knob to set the target time press the parameter setting knob to turn on the setting The LED time light will turn green then...

Page 9: ...n cannot display the actual temperature of sample in the vessel or dry bath surface temperature differences as following Dry bath center and outer edge The sample in the vessel and dry bath surface 7...

Page 10: ...alcohol Construction materials Water containing tenside Isopropyl alcohol Cosmetics Water containing tenside Isopropyl alcohol Foodstuffs Water containing tenside Fuels Water containing tenside Table...

Page 11: ...t NOTE This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection a...

Page 12: ...ensions W D H mm 175 290 85 Weight kg 1 6 without heating block Permissible ambient temperature 5 40 C Permissible relative humidity 80 Protection class according to DIN EN 60529 IP 21 Table 5 12 Orde...

Page 13: ...SCILOGEX LLC 1275 Cromwell Ave Suite C6 Rocky Hill CT 06067 USA Tel 1 860 436 9221 Fax 1 860 436 9745 info scilogex com www scilogex com...

Reviews: