II.

 

UNPACKING AND SET UP 

 

A.

 

Unpacking the Heating Base 

 

The Hybex Microsample Incubator heating base is shipped on its side within the interior 
carton. Remove any packing material around the unit and lift the unit out; taking care not to 
damage the hinged cover. Carefully inspect the heating unit for damage. Heat blocks and 
other accessories are packaged separately from the heating base.  

 

  

If there is evidence of damage, retain the shipping materials and contact SciGene 
(

[email protected]

) or your local distributor for assistance.

  

 

B.

 

Items Provided  

 

The following items are included with the Hybex heating base within the  
re-closable plastic bag: 

 

 

Power cord 

 

User Manual 

 

C.

 

Environmental Requirements 

 

Ensure that the area where the incubator is installed meets the following conditions, for 
reasons of safety and performance: 

Ambient tempe15 to +32°C (58 to 90°F) 

Relative humidity 15 to 80% RH non-condensing 

Air flow clearance 3 inches (8 cm) minimum on both sides 

 

 

D.

 

Installing the Instrument 

 

1.

 

Place the unit on a level surface within a few feet of the power source. Ensure the 
lid can be opened completely without interference and leave 3 inches of clearance 
along the back and sides for air circulation.  

2.

 

Plug the provided power cord into the back of the unit and then to a properly 
grounded outlet. Use only the power cord provided.  

 

Do NOT operate the system without a block inserted into the base unit! 

Ne pas faire fonctionner le système sans un bloc insérée dans l'unité de base! 

 
 

See 

Section VII. ORDERING INFORMATION

 for a list of available accessories and 

Section III. 

B. Handling Tube Blocks and the Waterbath Insert

 for how to insert and remove blocks. 

 

 

 

 

 

 

Allow 3 inches of 

clearance on both sides  

Allow 15 inches 
of total vertical 
clearance to fully 
open the lid. 

Allow 3 inches of 
clearance at the back 
for the power cord. 

Summary of Contents for 1057-30-0

Page 1: ...gene com 1287 Reamwood Avenue Sunnyvale CA 94089 USA Tel 408 733 7337 Fax 408 733 7336 techserv scigene com Hybex Microsample Incubator USER MANUAL Cat 1057 30 0 115V and 1057 30 2 230V FOR RESEARCH U...

Page 2: ...The foregoing warranty is void in the event the unit was abused or modified or used in a manner inconsistent with its intended purpose SciGene makes no other warranty expressed or implied including wa...

Page 3: ...ING YOUR HYBEX SYSTEM 7 A Components Controls and Accessories 7 B Handling Tube Blocks and the Waterbath Insert 8 C Using the Temperature Controller 8 D Calibrating the Temperature Controller 9 IV MAI...

Page 4: ...nseignements dans cette section concernant les risques et les dangers potentiels S assurer que toute personne impliqu e avec le fonctionnement de l instrument est instruit dans les pratiques g n rales...

Page 5: ...ator s Manual for further information and to proceed with caution ATTENTION Ce symbole vous avertit consulter ce guide de l utilisateur pour plus d informations et de proc der avec prudence D Warnings...

Page 6: ...of with standard waste at the end of its lifetime Instead it is your responsibility to return it to a designated collection point for electronic equipment recycling Contact your local SciGene distribu...

Page 7: ...llowing conditions for reasons of safety and performance Ambient temperature 15 to 32 C 58 to 90 F Relative humidity 15 to 80 RH non condensing Air flow clearance 3 inches 8 cm minimum on both sides D...

Page 8: ...olds tubes or plates for heating Microarray Incubation Chamber For array hybridization 2 needed for operation Waterbath Insert For heating buffers or water Digital Thermometer For calibrating system t...

Page 9: ...on d un insert de bain marie ne pas laisser le couvercle chauffant en position basse pendant plus de 90 minutes L exc s de la condensation se forme qui peut endommager la base chauffante C Using the T...

Page 10: ...thermometer Allow 1 minute for thermometer to stabilize The actual temperature of the block will be displayed 4 Calculate the difference between the digital thermometer and the temperature controller...

Page 11: ...ntation sur la position OFF et d branchez le cordon d alimentation avant d effectuer toute op ration de service Remove liquid or condensation wipe dry while instrument is OFF and block is cool or at r...

Page 12: ...rformance Temperature Range Ambient 5 C to 99 C Temperature Regulation 0 1 C Controller Controller Type Digital PID Display Four line LED Calibration Output T type Thermocouple Environmental Ambient t...

Page 13: ...set the controller to 65 C Allow 15 minutes to stabilize 2 Using the cable provided with the digital thermometer cat 1051 52 0 plug one end into the blue receptacle found on the back panel of the Hyb...

Page 14: ...evant standards shown below to indicate compliance EMC Directive 2014 30 EU Standards EN 61326 1 2013 Electrical equipment for measurement control and laboratory use to include LVD Directive 2014 35 E...

Reviews: