SciGene 1057-30-0 User Manual Download Page 11

 

Hybex

®

 Microsample Incubator — USER MANUAL

 

 

Version 1.

9

, Oct 2019                                                                      10                                            (408) 733-7337, [email protected] 

IV.

 

MAINTAINING YOUR HYBEX SYSTEM 

 

A.

 

Powering Off 

 

Turn the power switch to the OFF position and unplug the power cord before performing 
any service procedure. 

 

B.

 

Cleaning the Instrument 

 

Clean outer metal surfaces and the heated lid using a soft cloth and mild, detergent-based 
cleaner. Avoid abrasive cleaners that can scratch surfaces. Do NOT use caustic or strongly 
alkaline solutions (e.g., strong soaps, ammonia, or bleach).

 

 

 

If running radioactive or biohazardous reactions, consult your institution’s radiation safety 
officer or biosafety office for assistance.

 

 

Turn the power switch to the OFF position and unplug  
the power cord before performing any service procedure! 

 

Tournez le commutateur d'alimentation sur la position OFF et débranchez  
le cordon d'alimentation avant d'effectuer toute opération de service! 

 

Remove liquid or condensation (wipe dry) while instrument  
is OFF and block is cool or at room temperature! 

 

Enlever le liquide ou la condensation (essuyer), tandis que l'instrument  
est éteint et le bloc est frais ou à température ambiante! 

 

C.

 

Checking and Replacing Fuses 

There are two fuses located in a removable fuse block below the power cord receptacle on 
the back of the unit. A blown fuse appears dark. Always replace fuses with those of the 
same amperage and voltage as shown on the label below the fuse block. 

 

Turn the power switch to the OFF position and unplug  
the power cord before performing any service procedure. 

 

To remove fuses: 

1.

 

Insert a small, flat blade screwdriver 
into the tab recess just below the plug 
receptacle.  

2.

 

Push down to release the fuse block.  

3.

 

Slide fuse holder out from the fuse 
block while holding the retaining tab 
out of the way.  

4.

 

Gently pry out the fuses.  

 

Summary of Contents for 1057-30-0

Page 1: ...gene com 1287 Reamwood Avenue Sunnyvale CA 94089 USA Tel 408 733 7337 Fax 408 733 7336 techserv scigene com Hybex Microsample Incubator USER MANUAL Cat 1057 30 0 115V and 1057 30 2 230V FOR RESEARCH U...

Page 2: ...The foregoing warranty is void in the event the unit was abused or modified or used in a manner inconsistent with its intended purpose SciGene makes no other warranty expressed or implied including wa...

Page 3: ...ING YOUR HYBEX SYSTEM 7 A Components Controls and Accessories 7 B Handling Tube Blocks and the Waterbath Insert 8 C Using the Temperature Controller 8 D Calibrating the Temperature Controller 9 IV MAI...

Page 4: ...nseignements dans cette section concernant les risques et les dangers potentiels S assurer que toute personne impliqu e avec le fonctionnement de l instrument est instruit dans les pratiques g n rales...

Page 5: ...ator s Manual for further information and to proceed with caution ATTENTION Ce symbole vous avertit consulter ce guide de l utilisateur pour plus d informations et de proc der avec prudence D Warnings...

Page 6: ...of with standard waste at the end of its lifetime Instead it is your responsibility to return it to a designated collection point for electronic equipment recycling Contact your local SciGene distribu...

Page 7: ...llowing conditions for reasons of safety and performance Ambient temperature 15 to 32 C 58 to 90 F Relative humidity 15 to 80 RH non condensing Air flow clearance 3 inches 8 cm minimum on both sides D...

Page 8: ...olds tubes or plates for heating Microarray Incubation Chamber For array hybridization 2 needed for operation Waterbath Insert For heating buffers or water Digital Thermometer For calibrating system t...

Page 9: ...on d un insert de bain marie ne pas laisser le couvercle chauffant en position basse pendant plus de 90 minutes L exc s de la condensation se forme qui peut endommager la base chauffante C Using the T...

Page 10: ...thermometer Allow 1 minute for thermometer to stabilize The actual temperature of the block will be displayed 4 Calculate the difference between the digital thermometer and the temperature controller...

Page 11: ...ntation sur la position OFF et d branchez le cordon d alimentation avant d effectuer toute op ration de service Remove liquid or condensation wipe dry while instrument is OFF and block is cool or at r...

Page 12: ...rformance Temperature Range Ambient 5 C to 99 C Temperature Regulation 0 1 C Controller Controller Type Digital PID Display Four line LED Calibration Output T type Thermocouple Environmental Ambient t...

Page 13: ...set the controller to 65 C Allow 15 minutes to stabilize 2 Using the cable provided with the digital thermometer cat 1051 52 0 plug one end into the blue receptacle found on the back panel of the Hyb...