APPENDIX A – Using and Calibrating a Watlow SD31 Controller 

** For selected Hybex incubators manufactured between 2005 and 2010.** 

 

A.

 

Using the Watlow SD31 Temperature Controller 

 

The controller has a single LED panel with four buttons. The set 
temperature is displayed if the 

SET

 key is pressed and held otherwise the 

ACTUAL

 block temperature is displayed.  

 

To set the block temperature: 

 

1.

 

Hold in the 

SET

 button to view the previous set temperature.  

 

2.

 

While holding down the 

SET

 button, press the up or down arrow button until the 

desired set temperature is shown.  

 

3.

 

Release the buttons to display the actual temperature. The instrument will adjust 
block heating until the new temperature is attained. 

 

B.

 

Calibrating the Watlow SD31 Temperature Controller 

 

Re-calibrating the Hybex temperature controller is recommended 

only if

 the display varies 

by more than one degree (1°C) to a connected NIST-certified T-type digital thermometer 
(SciGene cat. #1051-52-0). 

 

 

To adjust the controller to achieve accurate temperatures: 

1.

 

Insert a block into the Hybex incubator, turn ON the instrument and set the 
controller to 65°C.  Allow 15 minutes to stabilize. 

2.

 

Using the cable provided with the digital thermometer (cat. #1051-52-0), plug one 
end into the blue receptacle found on the back panel of the Hybex incubator and 
the other end into the digital thermometer. 

3.

 

Turn on the thermometer. Allow 1 minute for thermometer to stabilize. The actual 
temperature of the block will be displayed. 

4.

 

Calculate the difference in the temperature shown on the controller and the digital 
thermometer. For example, if the controller displays 52.5°C and the digital 
thermometer displays 51.0°C, the difference (offset) is 1.5°C.  

5.

 

On the temperature controller, press the Infinity Key (

) for 

three seconds until “OPEN” appears.  

6.

 

Press the down arrow four times until “Cal” appears.  

7.

 

Press and hold the SET key. The existing offset value between 
the controller and digital thermometer is displayed.   

8.

 

Press and hold the SET key and use the up/down arrows to 
adjust the offset value to the temperature dHeaifference calculated in Step 4. For 
example, if the controller shows a temperature 1.5°C lower than the thermometer, 
adjust the offset by adding 1.5 to the value shown.  

9.

 

Press the Infinity Key (

) to exit calibra on and return to the opera on display. 

Verify that the temperature on the thermometer matches the display.  Your 
incubator is now calibrated to provide accurate temperatures between 35 and 99°C.

 

 

 

Summary of Contents for 1057-30-0

Page 1: ...gene com 1287 Reamwood Avenue Sunnyvale CA 94089 USA Tel 408 733 7337 Fax 408 733 7336 techserv scigene com Hybex Microsample Incubator USER MANUAL Cat 1057 30 0 115V and 1057 30 2 230V FOR RESEARCH U...

Page 2: ...The foregoing warranty is void in the event the unit was abused or modified or used in a manner inconsistent with its intended purpose SciGene makes no other warranty expressed or implied including wa...

Page 3: ...ING YOUR HYBEX SYSTEM 7 A Components Controls and Accessories 7 B Handling Tube Blocks and the Waterbath Insert 8 C Using the Temperature Controller 8 D Calibrating the Temperature Controller 9 IV MAI...

Page 4: ...nseignements dans cette section concernant les risques et les dangers potentiels S assurer que toute personne impliqu e avec le fonctionnement de l instrument est instruit dans les pratiques g n rales...

Page 5: ...ator s Manual for further information and to proceed with caution ATTENTION Ce symbole vous avertit consulter ce guide de l utilisateur pour plus d informations et de proc der avec prudence D Warnings...

Page 6: ...of with standard waste at the end of its lifetime Instead it is your responsibility to return it to a designated collection point for electronic equipment recycling Contact your local SciGene distribu...

Page 7: ...llowing conditions for reasons of safety and performance Ambient temperature 15 to 32 C 58 to 90 F Relative humidity 15 to 80 RH non condensing Air flow clearance 3 inches 8 cm minimum on both sides D...

Page 8: ...olds tubes or plates for heating Microarray Incubation Chamber For array hybridization 2 needed for operation Waterbath Insert For heating buffers or water Digital Thermometer For calibrating system t...

Page 9: ...on d un insert de bain marie ne pas laisser le couvercle chauffant en position basse pendant plus de 90 minutes L exc s de la condensation se forme qui peut endommager la base chauffante C Using the T...

Page 10: ...thermometer Allow 1 minute for thermometer to stabilize The actual temperature of the block will be displayed 4 Calculate the difference between the digital thermometer and the temperature controller...

Page 11: ...ntation sur la position OFF et d branchez le cordon d alimentation avant d effectuer toute op ration de service Remove liquid or condensation wipe dry while instrument is OFF and block is cool or at r...

Page 12: ...rformance Temperature Range Ambient 5 C to 99 C Temperature Regulation 0 1 C Controller Controller Type Digital PID Display Four line LED Calibration Output T type Thermocouple Environmental Ambient t...

Page 13: ...set the controller to 65 C Allow 15 minutes to stabilize 2 Using the cable provided with the digital thermometer cat 1051 52 0 plug one end into the blue receptacle found on the back panel of the Hyb...

Page 14: ...evant standards shown below to indicate compliance EMC Directive 2014 30 EU Standards EN 61326 1 2013 Electrical equipment for measurement control and laboratory use to include LVD Directive 2014 35 E...

Reviews: