Table 3-3 Data Analysis and Review (continued)

Corrective Action

Possible Cause

Symptom

Give all users read and write access to the folder and
then repeat the analysis. Refer to "Open a Plate" in
the 

Operator Guide

.

The user does not have read
and write access to the
C:\32Karat\ folder.

The DataReviewer
software opens and
then an "Analysis
failed" message opens.

From the 

Start

 menu in Windows Explorer, click 

All

Programs > Sciex PDF  > Preferences

. Make

The location for the output
folder in the Sciex PDF Writer
software is not correct.

The DataReviewer
software opens and
then an "Analysis
failed" message opens.

sure that the folder location is
C:\ProgramData\Sciex\DataReviewer\MethodReports
and then repeat the analysis.

Make sure that the Sciex PDF Writer software is
selected as the printer for the C100HT instrument in
the 32 Karat software. Refer to 

Configure the Controller

and the Software

.

The Sciex PDF Writer software
is not selected as the default
printer in the 32 Karat
software.

C100

HT

 Biologics Analyzer System

Installation Guide

27 / 31

RUO-IDV-09-3829-A

Troubleshooting

Summary of Contents for C100HT

Page 1: ...C100HT Biologics Analyzer System Installation Guide June 2018 RUO IDV 09 3829 A...

Page 2: ...hose manufacturers products as supplied by SCIEX for incorporation into its equipment and does not imply any right and or license to use or permit others to use such manufacturers and or their product...

Page 3: ...ID 14 Install the Cartridge in the System 15 Verify the Alignment of the Transport System 16 Configure the Photomultiplier Tube PMT Voltage 18 Inject a Sample and Determine the Detector Counts 18 Set...

Page 4: ...system Do not perform unauthorized procedures This guide is intended solely for a trained and qualified Field Service Employee FSE It contains information required to install and troubleshoot the C100...

Page 5: ...scription Item Service door 1 Cartridge door 2 Sample door 3 LED power indicator green when the system is on blue when it is off 4 Access cover 5 Right side cover 6 C100HT Biologics Analyzer System In...

Page 6: ...ronic Equipment label 3 No User serviceable Parts label 4 For Research Use only label 5 RoHS label 6 RS 232 connector 7 Voltage switch 8 Power switch mains supply connection and fuse holder 9 High Vol...

Page 7: ...erature for at least 20 minutes before use Figure 1 3 Cartridge in Cartridge Stand Description Item Gel reservoir 1 Cartridge purge port 2 Cartridge data key 3 Capillary windows 4 Cartridge stand 5 C1...

Page 8: ...oxic chemicals have been used with the system WARNING Environmental Hazard Do not dispose of system components in municipal waste Follow local regulations when disposing of components Determine which...

Page 9: ...f them as hazardous waste Comply with all of the local regulations for the storage handling and disposal of biohazardous toxic or radioactive materials C100HT Biologics Analyzer System Installation Gu...

Page 10: ...cartridge stand and the ship kit containing cables from the packaging foam 4 Remove the boxes containing the controller and monitor 5 For customers with the cart Remove the cart from the box and then...

Page 11: ...lack knob counterclockwise to remove the transport lock 3 Instruct the user to store the transport lock in a safe location for later use when moving the system C100HT Biologics Analyzer System Install...

Page 12: ...et c Connect the third power cable to the monitor and to an appropriately grounded line voltage outlet 4 For customers with the cart Connect the power cables a Connect the cart power cable to the cart...

Page 13: ...double click csmain exe b In the 32 Karat Software Enterprise window right click the DataAnalyzer icon and then click Open offline c Type C100HTSW in the User name field and C100HTPrivate in the Pass...

Page 14: ...figure 9 Click File Exit to close the 32 Karat software 10 In the 32 Karat Software Enterprise window click File Exit 11 Put the customer reference DVD in the DVD drive on the controller and then cop...

Page 15: ...ot dry out c Allow the cartridge to stand at room temperature for at least 20 minutes before use 2 Remove the cartridge from the cartridge stand 3 Remove the seal from the cartridge purge port and the...

Page 16: ...es the tray automatically moves to the park position CAUTION Potential System Damage Move the transport system one step at a time when performing the alignment If the vertical or horizontal position i...

Page 17: ...X M field 6 Verify the D position a Click D and then verify that the transport system moves to row D of the 96 well plate b Click Up and then verify that the capillary tips are centered in the sample...

Page 18: ...HT Glycan Labeling and Analysis buffer tray 1 Remove the buffer tray from the refrigerator and then allow it to stand at room temperature for at least 20 minutes before use 2 Remove the seal from the...

Page 19: ...has a raw count that is approximately 15 000 then set the photomultiplier voltage Refer to Set the Photomultiplier Voltage Set the Photomultiplier Voltage 1 Start the C100HT software and type Service...

Page 20: ...r to Test the Detector If at least one of the capillaries has a raw count that is approximately 15 000 then no changes are required Write down the voltage If there are no capillaries with a raw count...

Page 21: ...the user IDs and passwords and give them to the customer Table 2 1 Default Users in the C100HT Software Password User ID User Role Sciex100 Administrator Administrator SciexAdv100 SciexAdvanced Advanc...

Page 22: ...PNP0501 X Device Parameters where X is a number Make sure that the Everyone user has at least read permission for this entry If not the COM port scan procedure might fail if the user operating the C10...

Page 23: ...le door is open Cannot start a run The cartridge door is open Insert the data key The cartridge data key was not detected In the C100HT software click Disengage and then click Engage If the problem pe...

Page 24: ...locked purge filter Perform a filter check Refer to Test the Filter in the Operator Guide If the filter is blocked then replace it Refer to Replace the Purge Filter in the Operator Guide Centrifuge th...

Page 25: ...hen replace the cartridge Clean the probe aligner with a cotton swab moistened with DDI water Compare the peak intensity of the IST peak to the intensities of the DP2 and DP15 peaks from the The label...

Page 26: ...lytical grade water and then run the sample again with the same or a shorter injection time The sample concentration is too high The peaks are saturated Use new micro vials especially for steps relate...

Page 27: ...s Make The location for the output folder in the Sciex PDF Writer software is not correct The DataReviewer software opens and then an Analysis failed message opens sure that the folder location is C P...

Page 28: ...ode b In the Process Profile list click OQ Suitability Profile c Click Run Parameters and make sure that OQ Suitability Profile is selected in the Method list d Optional Click Sample Information and a...

Page 29: ...rt Log Files for Troubleshooting This option is only available to users with the Basic User or Advanced User role Note This option is also available to users with the Service User role C100HT Biologic...

Page 30: ...the previous 24 hours b In the Results table select the tests to be exported c In the Logging Information table select all of the items 5 Click OK The information is exported to a file in the Trouble...

Page 31: ...PN C15918 is required Quantity Description Part Number 1 12 well tube strip 5 pack C13796 1 96 well sample plate 609801 1 C100HT buffer tray 3 pack C13795 1 Calibration solution C13794 C100HT Biologi...

Reviews: