6.3 Removing and reinstalling the back panel
1.
Use the condenser service port to completely drain the condenser.
2.
Loosen the 2 screws in keyhole slots.
3.
Remove the 4 other screws.
4.
Lift the panel and use the steel cable to hang it in the service position.
5.
Disconnect the three tubes.
Figure 6.3-1
Remove X 4
Loosen X 2