1x carbonate hardness test or total hardness test
1x sticker for Service Pass.
1x reducer 1"– 3/4"
1x flush valve with flush hose
If part(s) of the delivery scope are missing, please contact your local SciCan Office
(see back of cover).
4.2 Assembly of the pressure vessel and the pressure vessel lid
•
Stand on the kick loops
9
with both feet.
•
Lift the pressure vessel
1
and turn it clockwise until the mantle
handles
11
are over the kick loops
9
.
•
Remove the transport protective cap
12
from the filter cartridge.
•
Check the O-ring seal of the filter cartridge
2
for correct seat in the
groove, dirt and damage.
Note:
The cartridge seat has been
lubricated with food-safe lubricant at the factory.
•
Stand on the kick loops
9
with both feet and place the pressure
vessel lid
3
on the pressure vessel lid
1
. The positioning of the arrow
marking on the lid handle
13
must line up with the “INSERT” groove.
•
Press the pressure vessel lid
3
down and turn clockwise until the lock engages
8
.
4.3 Assembly of inlet and outlet hoses
Note:
The inlet and outlet hoses are not included in the standard scope of delivery.
The use of the HYDROS hose sets is recommended.
•
Fit inlet hose
5
at the inlet of the connector head
4
and outlet hose
7
at the outlet
of the connector head
4
.
Note:
Inlet “IN” and outlet “OUT” of the connector head
4
are equipped with
O-rings as seals, therefore no additional flat seals may be used here. Make sure that
the O-rings are seated correctly.
!
Caution:
The max. tightening torque at the 1" and 3/4" connections must not exceed
15Nm! Only hose connections with flat seals may be used. Hoses with conical screw
connections damage the connections of the filter head and invalidate any guarantee
claims! Before assembly, note the direction of flow on the upper side of the filter head,
IN = water inlet, OUT = water outlet. Prior to installation, note installation dimensions
and operating position (Chapter 10). If no original hoses are used, the 1"- 3/4"
14
reducer
supplied must be used to ensure correct sealing of the return valve.
14