![background image](http://html1.mh-extra.com/html/scican/hydros/hydros_manual_1217214016.webp)
6.1 Removing the Filter Cartridge
•
Switch off the power supply of the HYDRIM unit (remove plug).
•
Close the inlet valve
6
at the inlet hose
5
.
•
Place the flush hose in a bucket and remove pressure from the filter system by
opening the flush valve
15
. Collect waste water in the 10L bucket.
Note:
If the escaping water is more than 1 litre, the inlet value
6
is not completely
closed or is blocked with scale.
•
Stand on the kick loops
9
with both feet while opening the pressure
vessel lid
3
by pressing the lock
8
and turning it counter-clockwise
as far as it will go.
•
Place the pressure vessel lid
3
vertically on the lid handle
13
.
•
Stand on the kick loops
9
with both feet while turning the pressure
vessel
1
counter-clockwise by the mantle handles
11
as far as it will go.
•
Take your feet off the kick loops
9
and press the pressure vessel
1
down with both
hands on the mantle handles
11
.
•
Remove used filter cartridge
2
from the pressure vessel
1
.
•
Place the used filter cartridge
2
in the sink with the connection facing down for
drainage (> 5 min).
•
Lock the used filter cartridge
2
with the transport protection cap
12
of the new filter
cartridge and discard according to the local regulations.
6.2 Inserting the Filter Cartridge
•
Check the O-ring seal of the new filter cartridge
2
for correct seat in
the groove, dirt and damage.
Note:
The cartridge seat has been
lubricated with food-safe lubricant at the factory.
•
Check the connector seat of the filter cartridge O-ring
2
in the
pressure vessel lid
3
for dirt and damage.
•
Place new filter cartridge
2
in the pressure vessel
1
.
•
Stand on the kick loops
9
with both feet; lift the pressure vessel
1
whole turning clockwise until the mantle handles
11
are over the kick
loops
9
.
•
Stand on the kick loop
9
with both feet and place the pressure
vessel lid
3
on the pressure vessel
1
. The positioning of the arrow
marking on the lid handle
13
must line up with the “INSERT” groove.
•
Press the pressure vessel lid
3
down and turn clockwise until the
lock engages
8
.
•
Switch on electrical power supply to the HYDRIM unit.
•
Execute the steps described under 5.3 to flush and bleed the new
filter cartridge
2
.
16