schwa-medico OPErATING MANuAL - EMP 4 ECO+
52
Program 9 + u9:
Anti-Adaptation – Dynamic Stimulation
Type of stimulation Dynamic stimulation (fast)
Parameter
Frequency: 80 Hz
Pulse width: 150 μs
Time: 20 min
Rising ramp: 0.5 s
Falling ramp: 0.5 s
Symbols on the
display during
stimulation
Rising phase
Falling phase
Description
The 4 channels are working with the same frequency
and pulse width.
phase 1
phase 2
build-up phase
run-down phase
working phase
pause
phase
pulse width
T = 1/f
}
channel 2
Channel 1
phase 1
phase 2
build-up
phase
run-down
phase
200 µs
T = 1/f
pause
phase
pulses
time
time
intensity
intensity
time
min
frequency
Hz
10
20
30
50
100
frequency
Hz
3
6
50
100
2
2
time
s
0,25
time
s
0,5
0,75
working
phase
working
phase
1
frequency
Hz
7,5
15
50
100
time
s
7
8
70
time
s
intensity
mA
2
time t
s
intensity I
mA
75 µs
}
}
time
intensity
build-up phase
run-down phase
run-down phase
working phase
pause
phase
7
8
70
time
intensity
2
Program 10 + u10:
Classic Gate Control 120 hz
Type of stimulation TENS
Parameter
Frequency: 120 Hz
Pulse width: 150 μs
Time: 20 min
Working phase: 0.25 s
Pause: 0.25 s
Symbol on the
display during
stimulation
Description
phase 1
phase 2
build-up phase
run-down phase
working phase
pause
phase
pulse width
T = 1/f
}
channel 2
Channel 1
phase 1
phase 2
build-up
phase
run-down
phase
200 µs
T = 1/f
pause
phase
pulses
time
time
intensity
intensity
time
min
frequency
Hz
10
20
30
50
100
frequency
Hz
3
6
50
100
2
2
time
s
0,25
time
s
0,5
0,75
working
phase
working
phase
1
frequency
Hz
7,5
15
50
100
time
s
7
8
70
time
s
intensity
mA
2
time t
s
intensity I
mA
75 µs
}
}
time
intensity
build-up phase
run-down phase
run-down phase
working phase
pause
phase
7
8
70
time
intensity
2