OPAL
PRODUCT LABEL
Symptoms
Possible causes
Measures
No function of the hydraulic
drive
Hydraulic drive defective
Couch in lowest position
Check if the couch can still be pumped up
properly or if it keeps collapsing.
Attention ! Danger of slipping !
Please watch out for leaking oil!
Contact the service department of Schupp (see
last page)
No function of the electric
drive
Activated motor protection due to
overheating
If the motor is operated for too long without
interruption (> 1 minute), the internal protective
device can switch off the drive when the couch is
under maximum load. Wait until the drive has
cooled down again.
Problem cannot be fixed
Unknown
Contact the Schupp service department (see last
page).
Be sure to pay attention to the following features
of your therapy couch:
Type, part number
Serial number
This data can be found on the type plate on the
frame of the foot section end of the couch.