background image

 

 

 

 

EN

 

INSTRUCTIONS FOR USE 

 

OPAL 

Therapy couches 

 

139435, 136435

 

 

 

 

[UDI-OPAL]

 

 

 
 

[Actual instruction manual.] 

 

 

Summary of Contents for OPAL 136435

Page 1: ...EN INSTRUCTIONS FOR USE OPAL Therapy couches 139435 136435 UDI OPAL Actual instruction manual ...

Page 2: ... storage and processing in electronic systems Infringements give rise to a claim for damages and may have penal consequences All rights reserved 2022 SCHUPP GmbH Co KG Warranty Liability These instructions for use are obligatory before carefully and completely before putting the product into operation For correct use of the medical device observe all instructions and safety notes The instructions ...

Page 3: ... 6 3 Height adjustment with electric motor 10 6 4 Adjustment of the headrest upholstery 1 3 parts 12 6 5 Adjustment of the arm part upholstery 3 parts 12 6 6 Adjustment of the roof kyphosis position 13 6 7 Adjustment of the footrest upholstery 13 6 8 Adjusting the chair position 14 6 9 Adjustment of the Trendelenburg position 14 6 10 Wheeled frame 15 6 11 Paper roll holder 15 6 12 Equipment option...

Page 4: ...apy tables from SCHUPP GmbH Co KG are developed in the Black Forest Baden Württemberg and manufactured with regional suppliers The assembly of the therapy tables takes place in our ISO certified production facility and thus guarantees a high and always consistent processing quality 2 0 Safety regulations 2 1 Safety signs Symbols Safety relevant sections instructions or notes are highlighted by the...

Page 5: ...6 42 EC Medical hydraulic equipment General requirements for safety 2 3 Intended use The therapy couch of the company Schupp GmbH Co KG is a medical device of class 1 EU MDR 2017 745 and is intended exclusively for professional use by medically trained personnel and instructed users for the proper positioning and storage of patients during therapeutic healing and medical treatments or for the shor...

Page 6: ...ed and authorised in work safety and accident prevention regulations to work on the therapy couch Caution The operator must ensure that all users have read and understood the instructions warnings and safety notes in these instructions and are sufficiently informed about their operation This must be checked at regular intervals The operator must ensure that the therapy couch cannot be operated by ...

Page 7: ...on is prohibited The stretcher has been designed exclusively for use under normal conditions indoors The therapy couch should be placed on a firm and level surface with adequate free space for the user Warning Observe the acclimatisation time before commissioning The power cable must be positioned in such a way that it does not get jammed between the moving parts of the couch and so that nobody ca...

Page 8: ...ssary safety inspections according to MPBetreibV Medical Devices Operator Ordinance BetrSichV Operational Safety Ordinance and DGUV German Social Accident Insurance 4 0 Warranty Guarantee Schupp GmbH Co KG guarantees you a therapy couch free of material and processing defects The couch was thoroughly tested before shipment and 100 approved in the final production test The warranty period is 60 mon...

Page 9: ...t of the couch frame In the case of an energetically operated therapy couch it should be positioned near a power socket Lay the mains cable in such a way that there is no danger use cable bridges if necessary Check whether the operating voltage of the therapy couch label corresponds to the mains voltage The therapy couch is immediately ready for operation after connection to the mains voltage 6 2 ...

Page 10: ...easing the button immediately interrupts the adjustment process Manual switch Foot switch Once the height adjustment has been completed within 3 seconds before the couch switches off automatically If the interruption is greater than 3 seconds you must execute the double pressure release again For wiring diagram see Technical Data The times are permanently coded in the drive and cannot be changed T...

Page 11: ...ess release again For switching diagram see technical data The therapy couch must not be operated by unauthorised persons when unattended Children must always be supervised and are not allowed in the area of the therapy couch Cleaning and care personnel must always be instructed in occupational safety DANGER Moving parts keep your distance when adjusting the position Never use the all round switch...

Page 12: ... section may be loaded with a maximum of 30kg Warning 6 5 Adjustment of the arm part upholstery 3 parts With the 3 piece headrest optional both armrests can be adjusted completely independently of each other The lowering of the armrest cushions is infinitely variable over a range of up to 18 cm The arm section can be fixed in the individual position by means of a star grip To adjust an armrest hol...

Page 13: ...ent with an all round handle Watch out for hands and objects when adjusting the body part pads Position adjustment inform patient beforehand and secure against slipping if necessary Warning 6 7 Adjustment of the footrest upholstery The adjustment of the footrest is damped by means of the gas spring in order to avoid sudden and rapid movements of the cushion You can raise or lower the footrest upho...

Page 14: ...t into deflection The headrest is adjusted by releasing the operating lever under the headrest cushion see chapter 5 4 The locking levers located on both sides next to the upholstery centre section allow the body part upholstery to be adjusted see chapter 5 6 The operating lever on the side of the footrest allows the footrest to be adjusted to the appropriate treatment position see chapter 5 7 Do ...

Page 15: ... foot The couch is lowered onto the feet in a cushioned manner and now stands firmly on the floor again The wheel frame is designed for moving the stretcher without a patient overload For stability treatments must only be carried out with the wheel frame inactive Never extend or retract the wheel frame while a patient is on the stretcher The shock generated during lowering can cause pain and injur...

Page 16: ... Pad width 65 cm Pad width 80 cm imitation leather 6cm durable composite structure imitation leather 8cm Comfort de Luxe padding extra soft Head section incl nose slot adjustable with GDF 30 30 Head section 3 tgl incl nose slot adjustable with GDF 30 30 arm sections adjustable up to 18cm Nose slit insert Headboard Pro Comfort one piece with 3 different inserts Inserts Lying surface 2 parts Lying s...

Page 17: ...tors foot plugs arm section guides as well as operating parts for the electric height adjustment 8 0 Service life maintenance and care The therapy couch you have purchased is designed for an extremely long service life The limiting factor here is the drive which is designed for a lifetime of max 10 years or 100000 cycles After this time depending on the condition of the table the drive can be repl...

Page 18: ...d or greased The piston rod of the drive should be regularly cleaned of dust with a soft cloth If the electric motor therapy couch is not used for a longer period of time it should be disconnected from the mains supply Power cable Check the power cord regularly If it is damaged it must be replaced immediately Please contact the Schupp service department for this purpose Wheeled frame To ensure smo...

Page 19: ...one etc leads to damage and after a short time to brittleness and discoloration of the artificial leather surface The upholstery must not be cleaned with alcoholic or chemical solvents such as thinner petrol or similar Do not use chlorides synthetic detergents wax polishes or aerosol sprays for cleaning Warning Always disconnect the mains plug before cleaning Electrical voltage never clean the the...

Page 20: ... Motor is not connected to the power supply Fuse is defective Cables are damaged Check that the couch is properly connected to a 230V socket and that this socket is live Check the fuse in the house distribution box Check the power cable for damage Check supply line of hand or foot switch for damage Check whether the connecting cables of the switching elements are correctly plugged into the motor E...

Page 21: ... internal protective device can switch off the drive when the couch is under maximum load Wait until the drive has cooled down again Problem cannot be fixed Unknown Contact the Schupp service department see last page Be sure to pay attention to the following features of your therapy couch Type part number Serial number This data can be found on the type plate on the frame of the foot section end o...

Page 22: ... waste must be recycled or disposed of in a manner that is safe for people and the environment Please inform yourself in principle about the regulations applicable in your country or also about the local regulations for the disposal of old devices and accessories Watch out for physical hazards such as sharp edges during disassembly disposal 12 0 Technical data Medical Devices smashing EU MDR Class...

Page 23: ...y adjustable and removable up to 18 cm Locking by means of star handles Body section upholstery 1 part Head section upholstery 3 parts Nose slit opening and cover Manually adjustable with gas pressure spring Adjustment range The 5 piece couch surface with roof position includes Armrest upholstery Arm rests independently adjustable and removable up to 18 cm Locking by means of star handles Body sec...

Page 24: ... Watt Duty cycle Short term operation Operating time 2 minutes operation 18 minutes break Service life max 100 000 cycles or 10 years Protection class Protective insulation class II Protection class Degree of protection IPx6 Low voltage LINAK CE Conformity as proof of compliance with the Low Voltage Directive 2006 95 EC according to the following standard EN 60601 1 2006 A1 CORR 1 CORR 2 EMC ESD L...

Page 25: ...ch off t3 Share Drive Timing diagram Time min s max s Explanation t1 0 3 1 5 t1 Switch on time for first second pressure switching unit t2 0 3 2 t2 Waiting time until second pressure switching unit t3 3 Waiting time until the enable is switched off If the maximum times are exceeded or t1 and t2 are not reached the release procedure at START starts again no release Switching element with single Man...

Page 26: ...220822 SCHUPP GmbH Co KG 72280 Dornstetten Aach Glattal Straße 78 T 49 0 7443 243 0 F 49 0 7443 243 255 info schupp gmbh de www schupp shop 26 OPAL DECLARATION OF CONFORMITY E ...

Page 27: ...220822 SCHUPP GmbH Co KG 72280 Dornstetten Aach Glattal Straße 78 T 49 0 7443 243 0 F 49 0 7443 243 255 info schupp gmbh de www schupp shop 27 OPAL DECLARATION OF CONFORMITY H ...

Page 28: ...ne 49 0 7443 243 399 Fax 49 0 7443 243 255 Email vertrieb schupp gmbh de Internet www schupp shop You can reach our CH REP under CHRN AR 20002650 DORER GmbH Lörracherstrasse 50 4125 Riehen Phone 41 0 79 6371094 Email Thomas dorer dorer gmbh com Visit us on site Schupp GmbH Co KG Glattal Street 78 72280 Dornstetten Head office Production of preparations and personal care products Schupp GmbH Co KG ...