Technical description
28
Technical description
Accessories
Accessory Length
of
cables
manufacturer, designation
Body sensor
1.0 m
Schulte-Elektronik GmbH,
USB-Transponder-Pad 2.2 m
Schulte-Elektronik GmbH
note:
The use of not listed or not released accessory (e.g. other producer’s electrodes / other
types) at the SISS BABYCONTROL monitors may cause reduced immunity or rather
enhanced electromagnetic emission.
The use of the accessory listed above is – as far as the function makes possible – also
practicable with SISS-observation-monitors of
SCHULTE-ELEKTRONIK GMBH, which will keep to the compliance with the corresponding
standard regarding immunity and emission behavior.
Concerning the USB -Transponder-Pad an analysis of risk has given, that there is no
influence on the essential performance criteria of SISS BABYCONTROL monitors. The
immunity against electromagnetic environmental phenomena was not tested.
The use of the accessory listed above with other electrical medical devices can lead to their
reduced immunity and their enhanced electromagnetic emission.