Installing and switching on the monitor
13
Installing and switching on the monitor
Place the monitor in a position where the displays are easily visible
on entering the room. With the optionally available multifunction case
the SISS BABYCONTROL is optimally protected.
Warning:
Do not place the SISS BABYCONTROL directly
over the bed. Otherwise the monitor could fall into
the bed and injure your child if the cable is pulled.
Before switching on the SISS BABYCONTROL, first apply the
respiration sensor to your child.
=> Brief Instructions: First Use Page 10
It is best to lead out the sensor cable at the foot end of the bed, so
that your child will not use it as a toy.
Switch on the monitor with the ON button
(6).
Warning:
Before switching on, check that the rotary switch is set
correctly for the alarm value to be monitored .
As soon as it is switched on, the monitor carries out a self-test: all
monitoring LED’s light up in turn. A short signal tone also sounds at
the beginning and end of the test.
This allows you to check the LED’s and the audible signal generator
for correct function each time you switch on the monitor.
The flashing green monitoring LED (3) together
with a soft clicking tone on the SISS
BABYCONTROL indicate regular respiration. You
can enable/disable the monitoring tone by pressing
the “Respiration monitoring tone” button
(8) .