J1034-UV-A/B/E DETECTOR Manual
5
MOUNTING AND ALIGNMENT
The Detector should be aligned with the help of the spirit level and the montage
screw and fixing screw. The sprit level is mounted inside the levelling plate. The
following picture illustrates the components to align the detector correctly. To
avoid measurement errors, the detector should be carefully levelled in the
horizontal plane.
The Detectorbase2 (105304) is mounted on the bottom of the UV detector. It
includes a sprit level which allows a quick alignment. It is routinely mounted to the
UV detector.
The additional available Detectorbase (105303) is a general purpose base plate
which fits directly to the J1033-Shadowring (103304). It allows a robust and quick
alignment to individual grounds.
Use the tilting screws to align the detector and control the correct settings with the
help of the spirit level. After that use the montage screws to fix the detector to the
ground.
Figure 3
UV detector mounted on Detectorbase (Art.No.: 105103). This part fits ideally to the
J1033-Shadowring, which can be seen on the next page. The UV-xx detectors may be directly
mounted on an individual base with two M4 screws (distance diameter 50 mm) without using the
Detectorbase2 (Art.No.: 105104).
SPIRIT LEVEL
TILTING SCREW
MONTAGE
SCREW
105103
105104