20
GB
Properties
Values
Electrical information
Operating voltage,
frequency
230 V version
220 ... 240 V ~ 50 /60 Hz
120 V version
100 V ~ 50 / 60 Hz and
120 V ~ 60 Hz
Power consumption, max.
W
300
Fuses, primary
230 V version
-
T 3,15H,250 V in accordance with
IEC 127-3/5
120 V version
-
M 6,3 A in accordance with UL198 G
Protection class
-
II
Overvoltage category
-
II
Lamp type
-
Halogen reflector lamp
Osram, type HLX 64653
Philips, type 13163
Lamp rated voltage
V
24
Lamp rated power
W
250
Average lamp lifetime
Level 4
h
1500
Level 5
h
150
Level 6
h
50
Lighting information
Maximum effective
light guide bundle diameter
mm
15
Total light flux at light guide exit
(Schott light guide, Ø 15 mm, typ. values)
Level 4
lm
500
Level
lm
1000
Level 6 (max. light flux)
lm
1300
Light entry angle (2
α
eff
)
Without supplementary optics
degrees
approx. 53
Heat protection filter
-
SCHOTT KG 2, 45 x 45
thickness = 2,0 mm, toughened
Approvals
-
230 V version
EMV, VDE
120 V version
cUL / UL
The KL 2500 LCD has been tested and certificated to the applicable standards on electrical
laboratory equipment (DIN EN 61010-1 and UL 3101-1) the 230 V version also conforms to
the standard on medical electrical equipment (DIN EN 60601-1). This enables manufactu-
rers to obtain easy approval with integration of the KL 2500 LCD into their medical products.
Die 230 V version features .
The right is reserved to make changes in the design and supplied items within the scope of on-going technical
development.
Summary of Contents for KL 2500 LCD
Page 2: ......