Operation
113
Setting up the
temperature
compensation TC
The calculation of the temperature compensation is based
on the preset reference temperature, Tref 20 or Tref 25 (see
chapter 4.6 C
ONFIGURATION
).
You can select one of the following temperature compensa-
tions:
l
Non-linear temperature compensation "nLF"
according to DIN 38404 or EN 27 888
l
linear temperature compensation "Lin" with a
coefficient that can be set in the range 0.001 ... 3.000 %/K
l
no temperature compensation
Note
ips
Select the following temperature compensations to work
with the test samples given in the table:
Application
notes
Test sample
Temperature
compensation TC
Display
indicator
Natural water
(ground water,
surface water,
drinking water)
nLF
according to DIN 38404
EN 27 888
Ultrapure water
nLF
according to DIN 38404
EN 27 888
Other aqueous
solutions
Set linear
temperature coefficient
0.001 ... 3.000 %/K
Salinity
(seawater)
Automatically nLF accord-
ing to IOT
Sal,