Table base
Operating table OPX mobilis and OPX mobilis RC –
E
DITION
05-03-GB
ID.-Nr.: 02008857
71
Under reservation of changes in construction and measurements.
Safety classification
2
Applied part
type B
According to European Standard 60601-2-46 the maximum admissible leakage currents meet the CF requirement.
Int 1min/4min
—
The appliance is not designed for uninterrupted duty.
A 4 minutes interval is recommended after 1 minute’s full-load operation.
Protection against
penetrating of fluids
IXP4
Protection against
inflammable
mixtures
AP
Fuses employed
—
Fine-wire fuses 5 ×20mm
F1 = 10A, slow-blow type / F2 = 2A, medium time lag type
Battery type
employed
—
4 pcs; 6 V; 13.5 Ah
Input AC of the
battery charger
—
117V or 230V / 50 to 60Hz / 0.4A
Output DC of the
battery charger
27.6V / 1.4A
Internal fuses of the
battery charger
—
Fine-wire fuses 5 × 20mm:
2 pcs 0.4A, slow-blow type / 1 pc 2.0A, slow-blow type
Manufacturer
Schmitz u. Söhne GmbH & Co. KG
P. O. Box 1461, D-58734 Wickede (Ruhr)
Zum Ostenfeld 29, D-58739 Wickede (Ruhr)
Germany
phone +49 (2377) 840
fax
+49 (2377) 84162
http://www.schmitz-soehne.de
e-mail Domestic sales dept.: [email protected]
e-mail Exports dept.:
e-mail After-sales service:
Measurements/weights/electrical data
OPX mobilis
200
300 C
300 CL
300 CE
300 CLE
RC30
RC30/G
RC30L
RC30L/G
RC40
RC40/G