5
2
Available Models
** Suitable for 95% of all applications. PA = Polyamide Monofilament.
If the material to be measured differs significant from the SCHMIDT calibration
material in diameter, rigidity, shape, etc., we recommend calibration using customer
supplied material. For this purpose a material sample of about 5 m should be supplied.
International unit of tensile force: 1 cN = 1.02 g = 0.01 N
Model
Tension Range
cN
**SCHMIDT Calibration
Material
Q-10
2 - 10
Filament: 25 tex
Q- 20
2 - 20
Filament: 25 tex
Q-30
3 - 30
Filament: 25 tex
Q-50
5 - 50
PA: 0.12 mm Ø
Q-100
10 - 100
PA: 0.12 mm Ø
Q-200
20 - 200
PA: 0.12 mm Ø
Q-300
20 - 300
PA: 0.20 mm Ø
Q-500
50 - 500
PA: 0.20 mm Ø
Q-1000
50 - 1000
PA: 0.30 mm Ø
The standard instruments can have the following modifications or changes:
- Deviating measuring ranges
- Material of the rollers.
- Adjustment to other materials.
The Operating Instructions can also be used for these instruments as their
handling is the same.
2.1 Specifications
Calibration:
According to SCHMIDT factory procedure
Accuracy:
± 1% full scale (FS) or ± 1 graduation on scale
Scale diameter:
51 mm
Temperature range: 10 - 45°C
Air humidity:
85% RH, max.
Housing material:
Chill-cast aluminium
Housing dimensions:
78 mm x 62 mm x 27 mm (L x W x H)
Weight, net (gross): approx. 300 g (400g)
Guide rollers:
V-grooved
Line Speed
max. m/min
Roller Material
Standard
1000
Hardcoated Aluminium
Code T
1000
Plastic (POM) black
Code W
1000
Nickel plated steel