Page 1.9
Chapter 1
Operating Elements
SCHILLER AT-2 6-Channel ECG Unit
SERVICE HANDBOOK Issue 2.1 November 1998
Manual Mode
Manual mode
provides a direct printout of the real-time ECG with full control of parameter
selection.
To start the manual recording of a real-time ECG, press the MANUAL Printout key
To stop the manual recording (printout) press the STOP key
The printout provides you with the following:
Six (selected) leads with lead identification.
On the lower edge, the chart speed, time and date, user identification and filter settings (if
on).
At the top, the heart rate as current average of 4 beats, trace sensitivity.
The following can be freely chosen during or before the recording:
Lead Group
by means of the LEAD FORWARD
and LEAD BACKWARD key
The following lead groups are selectable:
I, II, III
aVR, aVL, aVF
(Cabrera: aVL, I, -aVR / II, aVF, III)
V1, V2, V3 /
V4, V5, V6
II, aVF, III /
V2, V4, V5
V4, V5, V6 /
V7, V8, V9
Chart Speed
Select speed 5, 25 or 50mm/s by means of the SPEED keys:
Sensitivity
Select 5, 10 or 20 mm/mV by means of the SENSITIVITY keys:
Myogram Filter
Switch the filter ON or OFF with the FILTER key:
25 Hz or 35 Hz is displayed on the bottom line of the printout when
the filter is switched on.
Recentering
To re-centre the ECG traces, press the 1mV key