The unit does not respond when the ON/OFF
button is pressed.
No aerosol, or too little aerosol, is generated.
The fan is not working correctly.
If the unit does not operate as it should then please check the following:
Investigation of faults
Condition
Corrective action
Check that the power supply unit is correctly plug-
ged into the nebuliser and that the power cable
is correctly plugged into the nebuliser socket.
Check that the water reservoir has been filled to
the level indicated.
If you are using a rechargeable battery pack for
power supply then check that it is fully charged.
Also check that the battery pack has been cor-
rectly connected to the unit.
Check whether the medicine cup is damaged or if
a number of cups are inserted at the same time.
Check whether the ceramic oscillator has a build-
up of lime on it or is damaged.
Check that the fan wheel is sitting correctly and
securely on the motor shaft.