humimeter FS1 & FS1.1 Operating Manual
Page 28
10. Faults
If the measures listed below fail to remedy any faults or if the device has faults not
listed here, please contact Schaller GmbH.
Fault
Cause
Remedy
Measuring error
The temperature of the ma-
terial being measured is too
low or high. I.e. the material’s
temperature is lower than 0 °C
or higher than +40 °C.
The temperature of the
material being measured
has to be between 0 °C and
+40 °C.
Temperature discrepancy
between device and material
being measured
Let the temperature adjust
to the material being meas-
ured (permitted difference
of max. 3 °C).
Wrong product type
Check whether you have
selected the right product
type (product) before taking
a reading (see
Wrong filling quantity
Exactly fill in the filling
quantity displayed in the
product type name
(+/- 0.5 gram).
Mouldy or rain wet material
The accuracy decreases
significantly.
Stored and fermented corn
from whole grain silage
May lead to a too high dis-
played measuring value.
Frozen material or material
mixed with snow
The accuracy decreases
significantly.
Contaminated material
Highly contaminated mate-
rial such as long ears of
barley or foreign material
can strongly influence the
measuring result.