background image

Fiber Collimators Series 60FC

Adjustment_60FC    10.12.2020    © Schäfter+Kirchhoff GmbH

Schäfter+Kirchhoff GmbH    Kieler Str. 212    22525 Hamburg, Germany     +49 40 85 39 97-0    [email protected]    www.sukhamburg.com

1

Fiber Collimators Series 60FC

This document provides assistance in installing the fiber collimators series 60FC. It describes how a fiber cable 
is attached and how the collimation setting is adjusted.

1.  Before You Start

The fiber collimator is shipped with a protection cap for the fiber receptacle and with a front cap.

Please remove all the protection caps first and do not use them as beam 
dumps (risk of photo contamination).

Notice: 

There is a two-part protection cap on the receptacle. Please perform the following steps in order to remove 
this cap:

Figure   1: 

First, remove the threaded cap from 
the fiber receptacle of the fiber  
collimator.

Figure   2: 

Then, slightly loosen the small pin 
screw. Use the screwdriver 9D-12. 
Make sure to not loosen it too far, as it 
is small and easily lost.

Figure   3: 

Now, pull out the plug.

Summary of Contents for 60FC Series

Page 1: ...re You Start The fiber collimator is shipped with a protection cap for the fiber receptacle and with a front cap Please remove all the protection caps first and do not use them as beam dumps risk of p...

Page 2: ...with FC PC 0 polish connectors for fiber collimators with a coaxial axis 60FC 0 The fiber collimators are compatible to all fiber connectors type FC including fiber connectors with end caps Do not tou...

Page 3: ...nsitive fiber end face always insert the fiber connector s ferrule A at an angle with the connector key B properly aligned to the receptacle notch C Figure 5 When the ferrule tip is safely located in...

Page 4: ...crew to reduce the free play of the ferrule in the receptacle The free play in between the connector ferrule and receptacle is only a few microns but necessary for inserting the ferrule without force...

Page 5: ...is the optical wavelength and beam the collimated beam diameter 1 e2 level Refer to the laser instruction manual for all instructions regarding laser safety Do not stare directly into the laser beam w...

Page 6: ...info sukhamburg de www sukhamburg com 6 Figure 10 Insert the eccentric key into the large hole Now adjust for the focus setting by rotating the eccentric key Adjust the collimation by minimizing the...

Page 7: ...khamburg com 7 4 Adjustment tools For adjusting the fiber collimators series 60FC you need the following tools Figure 12 Screw driver type 9D 12 and eccentric key type 60EX 4 or type 60EX 5 Figure 13...

Page 8: ...obligation to pay compensation All rights in the event of granting of patents or registration of a utility model Sch fter Kirchhoff GmbH and the Sch fter Kirchhoff logo are registered trademarks The...