6.2.2.2 24H automatic monitoring
According to the specific requirements, after checking and adjusting the clock, set the
time and run mode, frequency weighting and time weighting used in statistics and
group name etc. In the measurement menu, when the cursor is moved to the bottom
line in blue background, change the second menu option to “24H” to go to the 24H
automatic measurement interface. In this case, the meter’s status display line displays
“Ready”. When the calendar clock reaches the integral o’clock, the meter automatically
starts the measurement. When its duration reaches the measurement time, the
measurement stops and the measurement in one period ends. The meter’s status
display line displays “Wait” till the next integral o’clock. It stops automatically till the
measurements in 24 time periods are completed. In the 24h automatic monitoring
process, the user can’t pause the operation. For the first period, the user can start the
measurement with the key. However, for the other periods, the meter automatically
starts the measurement when it reaches the same integral time point. If the user hopes
the instrument will continue the next 24H measurement automatically after a “24H”
measurement, the run mode of should be set as “Continuous”. Doing so, the instrument
will proceed the 24H measurement constantly until the data storage is fulfilled.
Note1: The other analyzers can't go started when the 24H mode is going on.
Note2: If there is a overload in the measurement, the "overload" indication will be
displayed for a while.
6.2.3 Integral Annalyzer
Pressing the key when the cursor is at "Measure" will enter the measuring interface.
In the integral analyzer, there are four mode: Sim., Big, List, Huge.