28
c) Limit
When the mode of "Limit" is chosen, shown below:
"Limit": When the value exceeds the setting limit, the instrument will be started. The
value can be chosen from 0 to 180.
When the cursor is moved to "Run mode", the user can enter setup interface of Interval
run mode by pressing or key.
d) Interval
When the mode of "Interval" is chosen, shown below:
"Inter.
△
T=1 min": the time between startup of two measurements, the user can choose
among 10s, 30s, 1min, 2min, 5min, 10min, 20min, 30min and 1hour. 1min means the
instrument will be started at the exact minute, 5min means the instrument will be
started at the exact 5 minute
Note1: The instrument will save the measurement data automatically after the
measurement is finished, during this period, the instrument won’t response to the order
of measurement startup, so the Ts should better shorter than
Δ
T.
Note2: When the setting measurement time Ts is longer than interval, the measure-
ment time will be Ts, and the interval will be prolonged.
7.2.3 Analyzer setup
Analyzer setup includes: Statistics, integral. Other options are valid only with authoriza-
tion.
a) Statistics
The setup of Ts and Run mode is seen in the previous section.
"Extra": close, the basic percentile levels are L5, L10, L50, L90 and L95. If the "Extra" is
open, besides the 5 basic ones, there are 7 percentile level including L1, L99 and five
user defined levels. And the measurement interface of Statistics will have one more
interface: EXTRA.