9
ZOC33
Section 2: Introduction
Thermal Control Unit (ZOCTCU)
An optional Thermal Control Unit (ZOCTCU) (Figure 2.5) is
available for applications where temperature swings my be
great enough to exceed the compensated range of the sen
-
sors. Exceeding the compensated temperature range can
induce errors in the pressure measurements. See Section
1,
Specifications
for more information on the compensated
temperature range. The ZOCTCU consists of a housing,
insulation and a proportional heater which will maintain
the temperature for the module at 40°C ±0.3°C.
Figure 2.5 - ZOC33 Thermal Control Unit (128 channel shown)
Summary of Contents for ZOC 33/64Px
Page 4: ...INTENTIONALLY LEFT BLANK...
Page 11: ...7 ZOC33 Section 2 Introduction Figure 2 3 Current ZOC33 64PxX2 Dimensions inches mm...
Page 15: ...11 ZOC33 Section 3 Installation Operation Figure 3 2 ZOC33 Analog Interface...
Page 20: ...16 ZOC33 Section 4 Electronic Components INTENTIONALLY LEFT BLANK...
Page 24: ...20 ZOC33 Section 4 Electronic Components Figure 4 5 ZOC33 Decoder Board Connector Pinouts...