Dräger Interlock 7000
31
Troubleshooting
5
Troubleshooting
Display
Cause
Remedy
Flow detected during Warm-Up
Breath sample has been blown too early.
Deliver breath sample when requested by the instrument.
Blow harder! / Repeat Test
Breath sample has been blown too soft.
Repeat test blowing harder.
Blow softer! / Repeat Test
Breath sample has been blown too hard.
Repeat test blowing softer.
Blow longer! / Repeat test
Breath sample has been blown too short.
Repeat test blowing longer.
Suck after Blow / Repeat test
Breath sample is not delivered correctly by
blowing and sucking.
Repeat test with blowing and immediately afterwards
sucking.
Suck longer! / Repeat Test
Breath sample has been sucked too short Repeat test sucking longer.
Check mouthpiece / Repeat test
Mouthpiece is not correctly inserted into
handset.
Insert mouthpiece correctly.
Please retry / Repeat test
During delivery of the breath sample a fault
has been detected.
Deliver breath sample again when requested by
the instrument.
Test failed
The measured breath alcohol concentration
is above the set limit value.
Repeat the test after waiting at least 15 minutes.
If the message appears again, the breath alcohol
concentration is above the limit value.
Low Temperature / Warm Up Handset
Temperature of the handset too low.
Detach handset and take it to a place with a moderate
temperature. Connect handset again and perform test.
High Temperature / Cool Down Handset
Temperature of handset too high.
Detach handset and take it to a place with a moderate
temperature. Connect handset again and perform test.
Service required in x days
Service interval will expire in x days.
Call an authorized Dräger Interlock service centre in
due time.
Service lockout
Service interval expired.
Call an authorized Dräger Interlock service centre.
Violation Reset
Too many violations have been detected.
Call an authorized Dräger Interlock service centre in
due time.