Transferring the patient from the radio-transparent stretcher 
 

  Once the patient has been transferred, it is necessary to lay it down to the support: bed, 

shower trolley, examination table …. 

 

 

 

 
Prior to the transfer, make sure:  

-  That  the  bed,  the  shower  trolley  ...  has  enough  space  under  its  base  to 
allow manipulation of the hoist without any risk.  
-  That  you  have  enough  space  around  the  patient  to  manipulate  half-
stretchers once the patient has been laid on the support  
- That the brakes the shower trolley , bed ... are blocked so that the support  
is stable and there is no risk of inadvertent movement 

 

 

1-  Bring  the  hoist  above  the  support  and  lift  down  the  arm  until  the  stretcher  lies  on  the 

support and the loops are loose. 
Make sure that the patient is in a proper position.  
 

2-  Remove the loops straps of the spreader bar, exit the hoist from the room in order to have 

more freedom of movement around the patient. 
 

3-  Unlock the stretcher 

 

4-  Break off the two halves of the stretcher. Putting the patient in a supine position, you will 

find it easier to remove the half-stretchers one after the other by placing your hands at the 
openings for better grip, and sliding half-stretchers. 
 

 

 
 
 
 
 
 
 

 

 

 
 
 
 
 
 
 

 
 
 

 
 

 

 

If it is necessary lateralize the patient alternately on both sides to remove the half-stretchers, 

then  turn  the  patient  on  the  back  or  in  any  position  that  secured  his  physiological  condition 
and / or devices, probes ... he could be equipped

 

 

 

Summary of Contents for S19 21 001

Page 1: ...DOMA031EN v1_06 2014 Radio transparent stretchers user manual Page 1 24 Radio transparent stretchers Maximum load 240 kg 300kg User manual...

Page 2: ...assembly 5 Using the stretcher 8 Supporting a lying patient 10 Lifting a patient lying on the ground 13 Transferring the patient 16 Technical specification 17 Cleaning and disinfection 18 Periodical i...

Page 3: ...21 002 depending on the model A set of four straps loops ref S82 99 003 A user manual ref DOMA031 This symbol is used to point out instructions and information related to work place safety where injur...

Page 4: ...turn at 360 around an axis In the case of a mobile spreader bar hoist equipped with an actuator it is recommended to use the stretcher with at least two caregivers one person will manipulate the hois...

Page 5: ...mful If a component is missing or deteriorated do not assemble and use the device contact your vendor or SCALEO Medical Using the stretcher requires a good experience of patient transfer and some trai...

Page 6: ...spreader bar Unpack the stretcher remove all the protection parts Lay down the parts of the stretcher on a flat area in order to facilitate the assembly The two operators position the two halves face...

Page 7: ...system Unlock the system slide the male part out of the female part Remove the male part of the stretcher first The patient may be lateralized to facilitate uncoupling Storage The stretcher assembly c...

Page 8: ...tible if necessary require proof of compatibility SCALEO Medical is not responsible for any incident following the use of an unsuitable combination Hoist for patients must NEVER be used to transfer li...

Page 9: ...tient monitoring one for handling the lift Make sure before any transfer that you have enough space on both sides of bed shower trolley to pass half stretcher Place a pillow under the patient s head t...

Page 10: ...es and postures 1 If the lateralization is possible a medical advice is mandatory lateralize the patient position a patient s leg above the other the corresponding arm over the other arm then rotate h...

Page 11: ...and center the spreader bar above the patient see above Move the spreader bar down to a distance sufficient so you can hang the loops Attach the four loops to the spread bar like shown above 5 Slowly...

Page 12: ...of the stretcher under the patient and if necessary lateralize him her Make sure before the transfer that you have enough space around the patient to pass half stretchers and maneuver the hoist Becaus...

Page 13: ...13 24 3 Slide the other half of the stretcher under the patient until it reaches the first half The second caregiver may maintain the patient in a lateral position so the assembly of the stretcher is...

Page 14: ...y measures to be taken 1 Move the hoist above the patient 2 Hang two loops of the stretcher to the spreader bar 3 Lift the stretcher at the head of the patient so as to pass the leg of the hoist under...

Page 15: ...necessary correct the position of the stretcher on the legs of the hoist then hang the two remaining loops to the spreader bar 7 Slowly lift the patient a few inches making sure that the patient is p...

Page 16: ...inadvertent movement 1 Bring the hoist above the support and lift down the arm until the stretcher lies on the support and the loops are loose Make sure that the patient is in a proper position 2 Rem...

Page 17: ...cation S1921001 maximum safety load 300kg all dimensions in mm S1921002 maximum safety load 240kg all dimensions in mm Main technical specification Classification Class I medical device 93 42 EEC dire...

Page 18: ...to avoid cross infection Prepare the equipment for disinfection Purpose Facilitate cleaning lower the level of contamination protect staff and environment Remove the slings sheets all the bed covers c...

Page 19: ...l devices once retired in clean places raised above ground and protected from dust moisture heat sunlight and risk of deterioration of packagings tears staples Product should be avoided for cleaning a...

Page 20: ...utting zone Loops Seams are not frayed The slings are not cut or worn Geometry The plastic plates are not bent unbalanced Put the stretcher on a flat surface the beams are right no abnormal deformatio...

Page 21: ...caused to operators patients or any third party in the following cases Non compliance of the instructions and recommendations supplied within the present User s Manual Using un adapted spare parts As...

Page 22: ...mean NB the stretcher may be separated in two parts therefore each half wears the same label including the serial number Pictogram symbol Meaning Manufacturer s name and adress Date of manufacturing P...

Page 23: ...De classe I suivant la r gle 1 de l annexe IX de directive europ enne 93 42 CEE Is compliant to est en conformit avec The council directive concerning medical devices 93 42 EEC By conformity to the Ha...

Page 24: ...ce This is a matter of responsibility for your establishment and a question of safety for your patients and staff Please do not hesitate to contact us if you require electrical diagrams lists of compo...