IPB50
16
dECI
(*) DECIMAL POINT POSITION
This step moves the position of the decimal point in the scale capacity value chosen from the previous step.
(For example, the scale capacity value “15000” would become 1500.0, 150.00, 15.000 depending on which of
the values: “
100000, 1.0, 1.00, or 1.000”
was selected respectively). The ->0<-s shown after the decimal
point indicate the number of digits after the decimal point. The value “
100000”
selects “no decimal point”.
(!)
1.000
ddES
(*) SET MINIMUM SCALE DIVISION
This is the minimum difference between two consecutive scale values. When it is a fraction of
Kg
, it can be
shown as
g
, tenth of a
g
or hundredth of a
g
by selecting one of the following value: 1, 2, 5, 10, 20, and 50.
With the ->T<- key select a value and confirm it with the
↵
key.
(!)
1
Note
: the maximum number of divisions that the instrument can display is 30.000. Therefore, if you insert a
value of maximum load, the minimum division to be insert has to be:
maximum load / minimum division
≤
30.000
If you insert a maximum load to 100000, the minimum division is 5 and if you try to insert 2, the instrument
goes in alarm and the display shows Err. Until you insert an acceptable value, you can not go ahead with the
set-up. By pressing
↵
, the instrument automatically inserts an acceptable value.
If you want to insert a different decimal point position, by pressing ->0<-, the display LED shows dECI then
press
↵
, set the decimal point position and go ahead with the configurations step.
0n-0
(*) AUTO->0<- AT SWITCH ON
Enables (
AUtO
) or disables (
FrEE
) the ability of the instrument to acquire a gross zero automatically when the
instrument is first switched on. When "
AUtO
" is enabled the display will be zeroing, if the gross weight value
on the scale is less than +/- 10% of the maximum capacity.
(!)
AUtO
tIM0
(*) ->0<- TRACKING TIME
The instrument periodically recalculates the zero when its variation does not exceed half the minimum scale
division. This step sets the length of time after which a new zero can be automatically taken. It is possible to
select values between 0 and 256 in increments of a tenth of a second.
(!)
10
In – Ir
(*) TILT DETECTOR ENABLING AND INFRARED REMOTE CONTROL
The functioning of this parameter varies depending on the version of hardware of the indicator. If an inclination
surveying device is connected, the tilt detector function can be enabled. In this case weighing is only possible
if the scale stays within 2% of horizontal (the maximum inclination from level as defined by law).
This parameter is also used to enable the infrared remote control, if an infrared sensor is present.
NonE : No selected functions
Inc 1: function currently not managed
Inc 2: function currently not managed
Ir 4:
function currently not managed
Ir 1:
function currently not managed
(!)
NonE
GAin
(*) AMPLIFIER GAIN SELECTION
Sometimes it may be necessary to change the amplifier gain. Although the standard value is "1.0," the user
can also select "0.5" or "1.5", “1.7” and “2”.
(!)
1