background image

Reference Sheet Version #005, 08-08-14; Firmware 2.4.2 and later 

Command List 

(commands most commonly used in field; see manual for complete list and detailed descriptions)

 

 

CATEGORY 

COMMAND 

DESCRIPTION 

MicroCAT 

ID 

ID? 

Get MicroCAT ID. 

*ID=ii 

Set MicroCAT ID to 

ii

, where ii= 0-99. Command must be sent twice. 

Global 

DateTime=x 

x=mmddyyyyhhmmss

 (set clock month, day, year, hour, minute, second.) 

GData 

Command 

all

 MicroCATs to run pump, take 1 sample, hold data in buffer until receiving 

Dataii

. Data not 

stored in FLASH memory. 

PwrOff 

Enter quiescent (sleep) state. Main power turned off, but data logging and memory retention unaffected. 

Get Data 

Dataii 

Get data obtained with 

GData

Status 

#iiGetCD 

Get and display configuration data. 

#iiGetSD 

Get and display status data. 

#iiGetCC 

Get and display calibration coefficients. 

#iiGetEC 

Get and display event counter data. 

#iiResetEC 

Reset event counter. 

#iiGetHD 

Get and display hardware data. 

#iiDS 

Get and display status. 

#iiDC 

Get and display calibration coefficients. 

General 

Setup 

#iiBaudRate=x 

x=

 baud rate (4800, 9600, 19200, 38400, 57600, or 115200). 

#iiRxDelay=x 

x=

 delay after MicroCAT receives command until transmitter enabled (1 - 500 msec). Default 25 msec.

 

#iiTxDelay=x 

x=

 delay after MicroCAT transmits reply until transmitter disabled (1 - 500 msec). Default 25 msec.

 

#iiDateTime=x 

x=mmddyyyyhhmmss

 (set clock month, day, year, hour, minute, second). 

#iiOutputExecutedTag=x  x=Y

: output XML Executed and Executing tags. 

x=N

: do not. 

#iiOutputBusyTag=x  x=Y

: output Busy tag if a command is sent while MicroCAT is sampling in response to 

GData

x=N

: do not.

 

#iiReferencePressure=x  x

= reference pressure (gauge) in db (used when MicroCAT has no pressure sensor). 

Pump  

Setup 

#iiMinCondFreq=x 

X

= minimum conductivity frequency (Hz) to enable pump turn-on for sampling. 

#iiAdaptivePumpControl=x

  x=Y

: Use 

Adaptive Pump Control

x=N

: Run pump for [

#iiOxNTau * #iiOxTau20

] before each sample. 

#iiOxNTau=x 

x

= pump time multiplier. Range 0 – 100.0;

 default 7.0

#iiPumpOn    #iiPumpOff 

Turn pump on/off for testing or to remove sediment. 

SBE 63  

Setup 

#iiSend63=command 

Command MicroCAT to send 

command

 to SBE 63, receive response (any command recognized by SBE 63). 

Other commands 

SBE 63

 setup required: 

SetEcho=1

SetFormat=1

SetAvg=1 to 16

 (recommend 

2

), 

SetAutoRun=0

Memory 

Setup 

#iiInitLogging 

Initialize logging to make entire memory available for recording. 

Output 

Format  

Setup 

#iiOutputFormat=x 

x=0

: raw decimal data. 

x=1

: converted decimal data.  

x=2

: converted decimal data in XML. 

#iiOutputTemp=x 

x=Y

: Output temperature. 

x=N

: Do not. 

#iiSetTempUnits=x 

x=0

: Temperature °C, ITS-90. 

x=1

: Temperature °F, ITS-90. 

#iiOutputCond=x 

x=Y

: Output conductivity. 

x=N

: Do not. 

#iiSetCondUnits=x 

Conductivity, specific conductivity units.                  

x=0

: S/m.                    

x=1

: mS/cm.                   

x=2

: µS/cm. 

#iiOutputPress=x 

x=Y

: Output pressure. 

x=N

: Do not. 

#iiSetPressUnits=x 

x=0

: Pressure decibars. 

x=1

: Pressure psi (gauge). 

#iiOutputOx=x 

x=Y

: Output oxygen. 

x=N

: Do not. 

#iiSetOxUnits=x 

x=0

: Oxygen ml/L. 

x=1

: Oxygen mg/L. 

#iiTimeFormat=x 

x=0 (

default

):

 dd mmm yyyy, hh:mm:ss. 

x=1:

 yyyy-mm-ddThh:mm:ss. 

x=2:

 hh:mm:ss, dd-mm-yyyy. 

x=3:

 hh:mm:ss, mm-dd-yyyy. 

#iiOutputSal=x 

x=Y

: calculate and output salinity (psu). 

x=N

: do not. 

#iiOutputSV=x 

x=Y

: calculate and output sound velocity (m/s). 

x=N

: do not.

 

#iiOutputSC=x 

x=Y

: calculate and output specific conductivity. 

x=N

: do not.

 

#iiTxSampleNum=x 

x=Y

: calculate and output sample number with data. 

x=N

: do not. 

Autonomous 

Sampling 

(Logging) 

#iiSampleInterval=x 

x

 = interval between samples (10-21600 sec). When commanded to start sampling with 

#iiStartNow

 or 

#iiStartLater

, MicroCAT runs pump, samples, stores data in memory, and goes to sleep at 

x

 sec intervals. 

#iiStartNow 

Start logging now. 

#iiStartDateTime=x 

x=mmddyyyyhhmmss

 (delayed logging start: month, day, year, hour, minute, second.) 

#iiStartLater 

Start logging at delayed logging start time. 

#iiStop 

Stop logging or waiting to start logging. Send 

#iiStop

 before uploading data. 

Polled 

Sampling  

(data not stored 

in FLASH 

memory unless 

noted; format 

specified by 

#iiOutput 

Format=

 

unless noted)

 

#iiTS 

Do not pump. Take sample, store in buffer, output data. 

#iiTSR 

Do not pump. Take sample, store in buffer, output data in raw decimal form. 

#iiTPS 

Run pump, take sample, store in buffer, output data. 

#iiTPSH 

Run pump, take sample, store in buffer, do not output data. 

#iiTPSS 

Run pump, take sample, store in buffer and in FLASH memory, output data. 

#iiTSN:x 

Do not pump. Take 

x

 samples, output data. 

#iiTPSN:x 

Run pump continuously while taking 

x

 samples and outputting data. 

#iiT63 

Do not pump. Take sample from SBE 63, output oxygen data in format set by 

SetFormat=

 in SBE 63. 

#iiSL 

Output last sample in buffer. 

#iiSLTP 

Output last sample in buffer, then run pump, take new sample, and store in buffer. 

#iiSLTPR 

Output last sample in buffer in raw decimal format, then run pump, take new sample, and store in buffer. 

#iiDNx 

Upload last 

x

 scans from FLASH memory. 

Data Upload 

#iiGetSamples:b,e 

Upload data from scan 

b

 to 

e

, in format defined by 

#iiOutputFormat

#iiDDb,e 

Upload data from scan 

b

 to 

e

, in converted decimal form (

#iiOutputFormat=1

) . 

 

Reviews: