13
Part No. 001300-EN, 07-m01-2022
Ceiling Lift Track Bridge Owner’s Manual
4. SPECIFICATIONS
Product
Product Weight
9.8 kg (21.6 lb)
Product Life
10 years
Safe Working Load (SWL)
272 kg (600 lb)
Water Ingress Protection
Rating of Device
IP21
Noise Level
Maximum 54 dBA
Medical Electrical
Equipment Class
Class 1
Protection Class
Type BF
Power Source
Input
100-240 VAC, 50-60Hz
Output
29.4 VDC, max 1A
Battery
Battery type: Lithium Ion, 25.2V, 2500mAh
Environmental Conditions
Operating Temperature
5 °C to 40 °C (41 °F to 104 °F)
Battery Charging
Temperature
10 °C to 40 °C (50 °F to 104 °F)
Operating Humidity
15 to 93%, non-condensing
Storage Temperature
-25 °C to 70 °C (-13 °F to 158 °F)
Storage Humidity
0 to 93%, non-condensing
Atmospheric Pressure
70 kPa to 106 kPA
WARNING: This equipment is not suitable in the presence of flammable
anesthetic mixtures with air or oxygen, or with nitrous oxide.
Controls
Control
Lift presence detected by magnetic sensor
Compliance
IEC 60601-1:2005 A1:2012+A2:2020 (Medical Electrical Equipment)
IEC 60601-1-11:2015 (Homecare)
ISO 10535:2006
CAN/CSA Z10535.1:15