                 TC-BA-e-1712 

The Power-  on/ Power-  off key  3-6 has to be pressed  and not released until `CAL` 
appears on the display. It lasts about 5 seconds from the last key operation until CAL 
will appear. 
The sensor 3-1 has to  be placed carefully onto the respective base material. Then 
the Zero key 3-3 has to be pressed, accompanied by a “beep” sound 
If now NF:H  (or  F:H)  is shown on the display, the  sensor has to be lifted for more 
than 5 cm. Then the Zero- key 3-3 has to be pressed and the instrument returns into 
measurement mode. Now factory settings have been finished.  
Comment: This procedure should always be done within 6 seconds. Otherwise it will 
be automatically cancelled and restoration is invalid. 

10  Notes  

 
The linearization of the instrument, which is given by the calibration, can be changed 
with the 

Ln- function.

 

It is strictly recommended not to perform any changes of the 

value of LN, 

as these 

changes can lead to deviant measurement results. 
 

Any adjustment or changed value of LN will seriously affect the accuracy. This 
value should only be adjusted by professional persons. 

 
Generally said: 
The bigger the value of Ln, the smaller the reading on the display for the same coat-
ing  thickness. Only a small change of the value of LN  causes a big change in the 
reading of the upper measurement range (at 500µm/20 mil). 
 
The 

value of LN

 has to be adjusted as follows: 

By pressing the Power-key: It lasts about 11 seconds from starting to press this key. 
This value can be changed by pressing the Plus- / Minus- key after `LN` appears on 
the display and the Power- key is released. This value is stored and afterwards the 
Zero- key has to be pressed.    
A: The reading value at low end has to be adjusted by pressing the Plus-/ or the Mi-
nus- key. 
B: The value of

 LN

 is enlarged if the reading value at low end (e.g. 51µm) is o.k., but 

the reading value at high end (e.g.432µm) is too big. 
In contrast with this, the value of LN has to be decreased, if the reading value at low 
end (e.g. 51µm) is o.k. but at high end (e.g. 432µm) it is too small. 
C: Procedures from A to B have to be repeated until the reading value for every cali-
bration foil is satisfactory in its accuracy. 
 
Annotation:  
To have a look at the CE Declaration of Conformity, please click onto the following 
link: 

https://www.kern-sohn.com/shop/de/DOWNLOADS/

 

 

Summary of Contents for TC 1250-0.1 F

Page 1: ...n E Mail info sauter eu Tel 49 0 7433 9933 199 Fax 49 0 7433 9933 149 Internet www sauter eu Instruction Manual Digital Coating Thickness Gauge SAUTER TC Version 1 2 08 2017 GB Illustration TC 1250 0 1FN TC BA e 1712 PROFESSIONAL MEASURING ...

Page 2: ...ge If you have any queries wishes or helpful suggestions do not hesitate to call our service number Models available TC 1250 0 1 F TC 1250 0 1 N TC 1250 0 1 FN TC_car 1250 0 1FN separate instruction manual available Summarize 1 Features 3 2 Specifications 3 3 Front Panel description 5 4 Measuring procedure 5 5 Zero Adjustment Calibration 6 6 Battery replacement 7 7 Adjustment foils 7 8 General adv...

Page 3: ...d on magnetic metals e g steel iron nickel etc The N sensor measures the thickness of non magnetic coatings on non magnetic metals It is used on anodizing varnish paint enamel plastic coatings powder etc It can be applied on non magnetic materials like aluminium brass stainless steel etc The N sensor measures the thickness of non magnetic coatings on non magnetic metals It is used on anodization v...

Page 4: ...V AAA UM 4 battery Operating conditions Temperature 0 to 50 C Humidity less than 80 Dimensions 127 x 67 x 28mm 5 0 x 2 6 x 1 1Inch Weight about 81g batteries not included Scope of delivery Carrying case Instruction manual According to types 1 Incorporated sensor at F model TC 1250 0 1 F with button F N 2 Incorporated sensor at N model TC 1250 0 1 N with button F N 3 Incorporated sensor at FN model...

Page 5: ...ymbol Fe F for magnetic metals or NFe N for non magnetic metals which is indicated on the display It switches to the automatic measurement mode which also correctly identifies the zeroing plate or any other base material b The sensor 3 1 has to be placed onto a coating layer to be measured The reading on the display is the thickness of the coating layer This can be corrected by pressing the Plus k...

Page 6: ...s the Power key 3 6 5 Zero Adjustment Calibration a Zero adjustment for Fe F and NFe should be carried out separately The Zero plate of iron has to be chosen if Fe is displayed and zero plate of alumi num has to be chosen if NFe is displayed Sensor 3 1 has to be placed carefully onto the appropriate base plate or another un coated base material The Zero key 3 3 has to be pressed without lifting th...

Page 7: ...ise a The instrument should always be calibrated on the uncoated base material to be measured instead of the base plate included in the delivery Then accuracy is more precise right from the start b Sensors will eventually wear off Life of the sensor will depend on the number of measurements taken and how abrasive the coating is The change of a sensor should only be performed by qualified persons 9...

Page 8: ...y be adjusted by professional persons Generally said The bigger the value of Ln the smaller the reading on the display for the same coat ing thickness Only a small change of the value of LN causes a big change in the reading of the upper measurement range at 500µm 20 mil The value of LN has to be adjusted as follows By pressing the Power key It lasts about 11 seconds from starting to press this ke...