Instruction Manual
FH
6. Configuration of RS 232
SUB-D 9pm
Pin
Signal
Illustration
2
TxD
Output signal
3
RxD
Input signal
5
GND
Ground
6
+1.6 to + 2 V
Over upper limit
7
+1.6 to + 2 V
Lower than lower limit
8
+1.6 to + 2 V
OK
6.1 Output Protocol
RS-232 Parameter
Baud rate:
9600
Data-Bit:
8
Parity:
none
Stop-Bit:
1
The measured value is requested by the PC by the ASCII
Sign “9”.
The measured value that comes from the instrument has
this format:
e.g.
0011.70
means -11,70 Newton, if
Newton is the selected unit
||-----|
|
|____>
the other 6 places describe the
measured value as ASCII-Signs
|_______>
the first place describes the
direction of the force (0 = minus = Pressure;
1 = plus = Tension)
or: 1021.15 means +21,15 N (Tension)
|_______> the first place describes the
direction of the force (0 = minus = Pressure;
1 = plus = Tension)
or: 1021.15 means +21,15 N (Tension)
7. Warning
Intended use
The instrument you have acquired serves to determine the
measuring value of the material to be measured. It is
intended to be used as a “non-automatic“ instrument, i.e.
the material to be measured is manually and carefully
attached at the instrument. The measuring value can be
read off after a stable measurement value has been
obtained.
Inappropriate use
Do not use the instrument for medical measurements. In
the event that small quantities are removed or added to the
material to be measured, incorrect measuring results can
be displayed due to the “stability compensation“ in the
instrument. (Example: Slow draining off of liquid from a
container suspended from the instrument). Do not attach a
continuous load. This can damage the measuring unit as
well as the parts, relevant to safety.
Prevent jolts, torsion and oscillation (e.g. by appending
slopingly) of all kinds. Be sure to prevent overloading the
instrument in excess of the stated maximum load (max.),
minus any tare weight that may possibly exist. This could
damage the instrument (risk of breakage).
Important:
• Always make sure that there are no people or materials
below the load that could be injured or damaged!
• The instrument is not suitable for measuring people. Do
not use as baby scales!
• The instrument does not comply with the medical product
law (MPG).
Never operate the instrument in hazardous locations. The
series design is not explosion-proof. Structural alterations
may not be made to the instrument. This can lead to
incorrect measuring results, faults concerning safety
regulations as well as to destruction of the instrument. The
instrument may only be used in compliance with the
described guidelines. Varying areas of application/ planned
use must be approved by SAUTER in writing.
Guarantee
The guarantee is not valid in following cases:
• Non-observation of our guidelines in the operating
instructions
• Use outside the described applications
• Alteration to or opening of the device
• Mechanical damage and damage caused by media,
liquids
• Natural wear and tear
• Inappropriate erection or electric installation
• Overloading of the measuring equipment
FH_SI-BA-e-1317