Place the decimal separa-

tor with the 

[ ][ ] 

and 

press the ENTER-button

ENTER:
Press ENTER to move to lower level in a menu or to accept a com-
mand or parameter value.
UP ARROW:
Use the UP arrow key to move up on the current menu level or to 
increase the selected parameter value. 
DOWN ARROW:
Use the DOWN arrow key to move down on the current menu level or 
to decrease the selected parameter value.
ESC:
Press the ESC to move back/higher in menu or cancel current action.

press the ESC- button to 

enter the menu

Basic settings 

select CONFIGURA-

TION and press the 

ENTER-button

press the ESC- button to 

enter the menu

2-calibration with water 
and one sample point

press the 

[   ] -

button and 

select MAOUTPUT and 

press the ENTER-button

select LRV (mA-output 

lower range value 4mA) 

and press the ENTER-but-

ton

insert lower range value 

(4mA) with the

[

g

g

g

][ ] 

and ENTER- buttons and 
press the ENTER-button, 

until upper separator 
reaches the right end 

of display. press the 

ENTER-button to store 

the lower range value for 

mA-output

press the 

[   ]

-button and 

select URV (mA-out upper 

range value 20mA) and 

press the ENTER-button

Place the decimal separa-

tor  with 

[   ]

 

[    

and 

press the ENTER-button

insert upper range value 

(20mA) with the

[ ][ ] 

and ENTER- buttons and 
press the ENTER-button, 

until upper separator 
reaches the right end 

of display. press the 

ENTER-button to store 

the upper range value for 

mA-output

press the 

[   ]

 -button 

and select DAMPING 

(time constant for mA-out 

damping) and press the 

ENTER-button

set the time constant with 

the 

[   ]

 

[   ]

-buttons 

and press the ENTER

press the ENTER-button 

to store the time constant 

for mA-output damping

press the  

[   ]

-button and 

select ALARMTYP and 

press the ENTER-button

Set alarm current with the    

[   ]

 

[   ]-

buttons (3.7 

or 22.5mA) and press the 

ENTER-button

press the ENTER-button 

to store the alarm current 

value

press the ESC-button 
to return to the main 

measuring screen

Collect sample

press the 

[   ] -

button and 

select NEWSAMPLE and 

press the ENTER-button

press the ENTER-button 

to activate sampling

the screeen will blink 

SAMPLING text during 

sampling process. Press 

the ENTER-button when 
sample has been taken to 

end sampling

the sample time stamp, 

average and min and max 
cs-values during the sam-

pling process are shwon on 

display.  Press ENTER to 
store the sample or press 

ESC to cancel

press the 

[   ] -

button and 

select SAMPLES and press 

the ENTER-button

p

p

p

p

p

Laboratory 

values

press the ESC- button to 

enter the menu

press the 

[   ] -

button and 

select CALIBRATION and 

press the ENTER-button

select witht he 

[   ]

 

[   ] 

buttons the desired sample 

point to which laboratory 

value will be inserted  and 

press the ENTER-button

Place the decimal separa-

tor  with 

[   ]

 

[    

and 

press the ENTER button

insert the laboratory value 

with the  

[   ]

 

y

 ] 

and 

ENTER- buttons and press 

the ENTER-button, until 

upper separator reaches 

the right end of display

press the ESC-button 
to return to the main 

measuring screen

Start-up cali-

bration

press the ESC- button to 

enter the menu

press the 

[   ] -

button and 

select CALIBRATION and 

press the ENTER-button

press the 

[   ] -

button and 

select CALIBRATE and 

press the ENTER-button

press the 

[   ]

-button and 

select the SAMPLEH2O 

to be the 1st calibration 

point and press the 

ENTER-button

press the

[   ] 

button 

and select the second 

point for calibration 

(SAMPL 01..09) and press 

ENTER-button.the display 

rolls the new calculated 

OFFSET, GAIN values. 

press ENTER to store val-

ues or press ESC to cacel

press ESC-button to return  

to measuring screen

1-Point         

calibration

press the ESC- button to 

enter the menu

press the 

[   ] -

button and 

select CALIBRATION and 

press the ENTER-button

press the 

[   ] -

button and 

select CALIBRATE and 

press the ENTER-button

press the 

[   ]

-button 

and select the sample 

point (SAMPL 01...09) for 

1-point calibration and 

press the ENTER-button

press ESC-button (1-point 

calibration, no second 

point). the display rolls the 

new calculated OFFSET, 

GAIN values. press 

ENTER-button to store 

ESC-button to cacel.

press ESC-button to return 

to the main measuring 

screen

2-point calibration with 

two sample points

2-Point           

calibration

press the ESC- button to 

enter the menu

press the 

[   ] -

button and 

select CALIBRATION and 

press the ENTER-button

press the 

[   ] -

button and 

select CALIBRATE and 

press the ENTER-button

select with the 

[ ][ ] 

-buttons the fi rst sample 

point for calibration  

(SAMPL 01...09) and press 

the ENTER-button

select with the 

[   ][   

-buttons the second 

sample point for calibra-

tion (SAMPL 01...09) and 

press the ENTER-button. 

the display rolls the new 

calculated OFFSET, GAIN 

values. press ENTER-but-

ton to store ESC-button 

to cacel.

press ESC-button to return  

to measuring screen

select RECIPE and press 

the ENTER-button

OFFSET           

adjustment

press the ESC- button to 

enter the menu

press the 

[   ] -

button and 

select CALIBRATION and 

press the ENTER-button

select OFFSET and press 

the ENTER-button

place the decimal sepa-

rator with the 

[   ]

 

p

 ] 

 

buttons and press ENTER

insert the OFFSET value 

with the  

[   ]

 

[   ] 

and 

ENTER- buttons and press 

the ENTER-button, until 

upper separator reaches 

the right end of display

press the ENTER-button 

to store the OFFSET value

press ESC-button to return  

to main measuring screen

select RECIPE and press 

the ENTER-button

GAIN            

adjustment

press the ESC- button to 

enter the menu

press the 

[   ] -

button and 

select CALIBRATION and 

press the ENTER-button

place the decimal sepa-

rator with the 

[   ]

 

p

 ] 

 

buttons and the ENTER

insert the GAIN value 

with the  

[   ]

 

[   ] 

and 

ENTER- buttons and press 

the ENTER-button, until 

upper separator reaches 

the right end of display

press the ENTER-button 

to store the GAIN value

press ESC-button to return  

to main measuring screen

press the 

[   ] -

button 

and select GAIN and press 

the ENTER-button

press the ESC- button to 

enter the menu

TIME and 

DATE settings

  

select CONFIGURA-

TION and press the 

ENTER-button

press the 

[   ] -

button 

and select SYSTCONF and 

press the ENTER-button

press the 

[   ] -

button and 

select SETCLOCK and 

press the ENTER-button

insert date with 

[ ][   ] 

-buttons (dd.mm.yyyy), 

press ENTER-button to 

move from dd->mm-

>yyyy and press the 

ENTER-button

insert time with 

[   ][   ] 

-buttons (hh.mm.ss), press 

ENTER-button to move 

hh->mm->ss and press the 

ENTER-button

press ESC-button to return  

to main measuring screen