SATO SK-8300 Instruction Manual Download Page 14

12 

 

Maintenance 

The measuring accuracy could be compromised if the lens is not clean. 

Use  a  camera  blower  to  clean  the  lens.  If  it  is  still  dirty,  use  a  cotton  swap  soaked  in  a  lens  cleaning 

solution to wipe the stain away. 

Caution:

 Never use  water  or any other cleaning solvent to clean the  lens. Doing so could damage the 

lens, causing the efficiency of the lens and measurement accuracy to fall. 

 

 

Error Messages 

Message 

Cause 

Solution 

 

Measured value is higher 
than 1550°C. 

Use the unit within the measuring range. 
If doing so still does not solve the problem, there is a 
possibility that the unit has other issues.  
Stop using the unit and consult us or the dealer from which 
the unit was purchased for assistance. 

 

Measured value is lower 
than -30°C. 

 

 

Trouble Shooting 

Trouble 

Possible cause 

Solution 

The power cannot be turned on. 

The battery is exhausted. 

Replace with a new battery. 

Laser marker does not work. 

The laser marker setting 
is OFF. 

Switch the laser marker setting to ON. 

Abnormal value is displayed. 

Measurement area does 
not suit the measurement 
object. 

Check the relationship between 
measurement distance and 
measurement area. 

There was a change in 
temperature when 
measuring the object. 

Check the measured value after taking 
measurements of objects with stable 
temperature. 

Dust or a stain was on the 
lens. 

Clean the lens. 

There was an extreme 
temperature change in the 
ambient. 

Let the instrument adapt to the 
ambient before use. 

Emissivity setting was 
inappropriate. 

Use a correct emissivity. 

 

If the problem is not solved, there is a possibility that the unit is defective. 

Please contact us or the dealer from which the unit was purchased for assistance. 

Summary of Contents for SK-8300

Page 1: ...nfrared Thermometer Model SK 8300 The SK 8300 complies with Portable Laser Applied Device of Consumer Product Safety Act and carries PSC mark 3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan URL http...

Page 2: ......

Page 3: ...asuring error or failure may occur resulting in a critical damage Never disassemble or modify the unit or use different types of batteries other than indicated as these actions may result in measuring...

Page 4: ...the unit was purchased Overview The SK 8300 is non contact type infrared thermometer which detects the infrared energy emitted from an object and converts it into a temperature reading The product can...

Page 5: ...measurement and alarm 6 Loop for wrist strap 7 Laser emitting window 8 Temperature measurement section lens 9 Battery compartment button Push this button to open the battery compartment cover 10 Meas...

Page 6: ...a stored data 8 Light up during T mode 9 Light up while setting emissivity 10 HAL Light up while setting upper threshold for alarm 11 LAL Light up while setting lower threshold for alarm 12 AUTO Ligh...

Page 7: ...r is pulled Open the battery compartment cover by pressing the battery compartment button Set the laser switch to ON The factory set is to ON Reinstall the cover and push in to fasten with a click The...

Page 8: ...ed mode at the time of turning the power OFF Normal Mode Measurement mode at normal MAX Mode To display a maximum value MIN Mode To display a maximum value AVG Mode To display an average value T Mode...

Page 9: ...gories Normal MAX MIN AVG or T Note Auto Measurement cannot be set during a Setting mode 3 Press the key AUTO is displayed on the LCD and a continuous measurement starts SCAN character blinks during a...

Page 10: ...Mode A measured value is displayed on the larger measurement displaying area and a setting value is displayed on the smaller one All setting values are stored even the battery is taken out DATA displa...

Page 11: ...isplay area is stored About Data Numbers Select a data number from 1 to 9 and A where the data is to be stored by using and keys The data numbering sequence is as follows DATA1 DATA2 DATA9 DATA A If a...

Page 12: ...he temperature measurements of unwanted parts resulting incorrect readings Emissivity All objects emit infrared radiation in accordance to surface temperature but the emissivity value can differ depen...

Page 13: ...eet the temperature value read from the contact type thermometer The emissivity found from this procedure could be used as a standard emissivity for the object Cautions on Measurement Measuring throug...

Page 14: ...lower than 30 C Trouble Shooting Trouble Possible cause Solution The power cannot be turned on The battery is exhausted Replace with a new battery Laser marker does not work The laser marker setting i...

Page 15: ...an alkali battery Backlight Standard Dimensions 51 W x 166 D x 200 H mm Weight approx 280 g with battery Materials Unit ABS resin Standard accessories Carrying case 9V battery Hand strap Warranty Poli...

Page 16: ......

Reviews: