SATO SK-8300 Instruction Manual Download Page 10

 

 AVG Mode:

 

 

The  “AVG”  character  lights  up  and  an  average  value  is  displayed  on  the  smaller  measurement 

displaying area. An average value is a mean temperature value since the AVG Mode is turned on. 

              Note:

  The  data  that  can  be  used  to  calculate  the  average  is  the  maximum  256.    When  the 

number of data exceeds 256, the average value is calculated again from them. 

 ΔT Mode:

 

 

The “ΔT” character lights up and the “ΔT” value is displayed on the smaller measurement displaying 

area.  A “ΔT” value is a difference between the maximum and minimum temperature values among 

the measured data after switching to the ΔT Mode. 

 

Caution

: Once  a  value  is  held  using  the  HOLD  function  and  the  trigger  is  pulled  to  retake  a 

measurement, all values, MAX, MIN, AVG and ΔT, are reset and displays a new set of  values 

for all modes. Also all values are reset once the power is turned off. 

 

Setting Mode 

A  measured  value  is  displayed  on  the  larger  measurement  displaying  area  and  a  setting  value  is 

displayed on the smaller one. All setting values are stored even the battery is taken out. 

 

∙ DATA display 

The “DATA” character lights up and a stored value is loaded on the smaller measurement 

displaying area. Refer to “Storing Data Function” for details. 

∙ 

ε

 

:

 

This enables the emissivity setting.  

 

 

Use 

 and 

 keys to set a emissivity value. 

 

  Refer to “About Emissivity” for more information. 

∙ HAL: 

Set an upper threshold for alarm. 

 

 

Use 

 and 

 keys  to  set  an  upper  threshold  value.  When  a  measured  value  exceeds 

the set value, the alarm sounds for alert. 

∙ LAL:

Set a lower threshold for alarm. 

 

 

Use 

 and 

 keys to set a lower threshold value.  When a measured value undergoes 

the set value, the alarm sounds for alert. 

 

* About Setting Alarm Sound 

   It is possible to set the alarm sound to be ON or OFF for both upper and lower thresholds. 

1.  Select either HAL or LAL in the Setting Mode. 

2.  Switch the alarm sound ON or OFF by pressing the   key. 

When the alarm sound is set to ON, the 

 lights up. 

Summary of Contents for SK-8300

Page 1: ...nfrared Thermometer Model SK 8300 The SK 8300 complies with Portable Laser Applied Device of Consumer Product Safety Act and carries PSC mark 3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan URL http...

Page 2: ......

Page 3: ...asuring error or failure may occur resulting in a critical damage Never disassemble or modify the unit or use different types of batteries other than indicated as these actions may result in measuring...

Page 4: ...the unit was purchased Overview The SK 8300 is non contact type infrared thermometer which detects the infrared energy emitted from an object and converts it into a temperature reading The product can...

Page 5: ...measurement and alarm 6 Loop for wrist strap 7 Laser emitting window 8 Temperature measurement section lens 9 Battery compartment button Push this button to open the battery compartment cover 10 Meas...

Page 6: ...a stored data 8 Light up during T mode 9 Light up while setting emissivity 10 HAL Light up while setting upper threshold for alarm 11 LAL Light up while setting lower threshold for alarm 12 AUTO Ligh...

Page 7: ...r is pulled Open the battery compartment cover by pressing the battery compartment button Set the laser switch to ON The factory set is to ON Reinstall the cover and push in to fasten with a click The...

Page 8: ...ed mode at the time of turning the power OFF Normal Mode Measurement mode at normal MAX Mode To display a maximum value MIN Mode To display a maximum value AVG Mode To display an average value T Mode...

Page 9: ...gories Normal MAX MIN AVG or T Note Auto Measurement cannot be set during a Setting mode 3 Press the key AUTO is displayed on the LCD and a continuous measurement starts SCAN character blinks during a...

Page 10: ...Mode A measured value is displayed on the larger measurement displaying area and a setting value is displayed on the smaller one All setting values are stored even the battery is taken out DATA displa...

Page 11: ...isplay area is stored About Data Numbers Select a data number from 1 to 9 and A where the data is to be stored by using and keys The data numbering sequence is as follows DATA1 DATA2 DATA9 DATA A If a...

Page 12: ...he temperature measurements of unwanted parts resulting incorrect readings Emissivity All objects emit infrared radiation in accordance to surface temperature but the emissivity value can differ depen...

Page 13: ...eet the temperature value read from the contact type thermometer The emissivity found from this procedure could be used as a standard emissivity for the object Cautions on Measurement Measuring throug...

Page 14: ...lower than 30 C Trouble Shooting Trouble Possible cause Solution The power cannot be turned on The battery is exhausted Replace with a new battery Laser marker does not work The laser marker setting i...

Page 15: ...an alkali battery Backlight Standard Dimensions 51 W x 166 D x 200 H mm Weight approx 280 g with battery Materials Unit ABS resin Standard accessories Carrying case 9V battery Hand strap Warranty Poli...

Page 16: ......

Reviews: