SATO SK-8300 Instruction Manual Download Page 1

 

 

 

 

 

Instruction Manual  

No.8268-00 Infrared Thermometer  

Model SK-8300 

 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 

The SK-8300 complies with Portable Laser Applied Device of Consumer Product Safety Act and 
carries PSC mark. 

 
 

 
 

3-4, Kanda-kajicho, Chiyoda-ku, Tokyo 101-0045 Japan 

URL: http://www.sksato.co.jp/english/ 

N.10 
 

Summary of Contents for SK-8300

Page 1: ...nfrared Thermometer Model SK 8300 The SK 8300 complies with Portable Laser Applied Device of Consumer Product Safety Act and carries PSC mark 3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan URL http...

Page 2: ......

Page 3: ...asuring error or failure may occur resulting in a critical damage Never disassemble or modify the unit or use different types of batteries other than indicated as these actions may result in measuring...

Page 4: ...the unit was purchased Overview The SK 8300 is non contact type infrared thermometer which detects the infrared energy emitted from an object and converts it into a temperature reading The product can...

Page 5: ...measurement and alarm 6 Loop for wrist strap 7 Laser emitting window 8 Temperature measurement section lens 9 Battery compartment button Push this button to open the battery compartment cover 10 Meas...

Page 6: ...a stored data 8 Light up during T mode 9 Light up while setting emissivity 10 HAL Light up while setting upper threshold for alarm 11 LAL Light up while setting lower threshold for alarm 12 AUTO Ligh...

Page 7: ...r is pulled Open the battery compartment cover by pressing the battery compartment button Set the laser switch to ON The factory set is to ON Reinstall the cover and push in to fasten with a click The...

Page 8: ...ed mode at the time of turning the power OFF Normal Mode Measurement mode at normal MAX Mode To display a maximum value MIN Mode To display a maximum value AVG Mode To display an average value T Mode...

Page 9: ...gories Normal MAX MIN AVG or T Note Auto Measurement cannot be set during a Setting mode 3 Press the key AUTO is displayed on the LCD and a continuous measurement starts SCAN character blinks during a...

Page 10: ...Mode A measured value is displayed on the larger measurement displaying area and a setting value is displayed on the smaller one All setting values are stored even the battery is taken out DATA displa...

Page 11: ...isplay area is stored About Data Numbers Select a data number from 1 to 9 and A where the data is to be stored by using and keys The data numbering sequence is as follows DATA1 DATA2 DATA9 DATA A If a...

Page 12: ...he temperature measurements of unwanted parts resulting incorrect readings Emissivity All objects emit infrared radiation in accordance to surface temperature but the emissivity value can differ depen...

Page 13: ...eet the temperature value read from the contact type thermometer The emissivity found from this procedure could be used as a standard emissivity for the object Cautions on Measurement Measuring throug...

Page 14: ...lower than 30 C Trouble Shooting Trouble Possible cause Solution The power cannot be turned on The battery is exhausted Replace with a new battery Laser marker does not work The laser marker setting i...

Page 15: ...an alkali battery Backlight Standard Dimensions 51 W x 166 D x 200 H mm Weight approx 280 g with battery Materials Unit ABS resin Standard accessories Carrying case 9V battery Hand strap Warranty Poli...