background image

 

 

 

For the model SK-101R (w/ auto temperature correction) 

Accessories: Wipe-out cloth, carrying case, screwdriver, dropper, saturation salt water 

 

- Using the Refractometer after Adjustment with Saturation Salt Water - 

 

Note: Be sure to perform the scale adjustment before using this instrument for the first time each 

day, or if the ambient temperature changes. 

 
(1)    Place yourself facing the light source and look through the eyepiece to see the scale. Adjust 

the focus to suit your eyes by rotating the eyepiece clockwise or counterclockwise. 

 
(2)    While holding the instrument as shown in Figure 1, open the cover plate, place a few drops 

of  saturation  salt  water  on  the  prism  and  then  close  the  cover  plate.  Look  through  the 
eyepiece to see the white and blue fields as shown in Figure 2 below. 

 
(3)    Scale adjustment 

This  instrument  is  designed  to  allow  for  adjustment  at  20°C,  and  the  intersection  of  the 
boundary  of  the  white  and  blue  fields  is  set  to  29.6%.when  the  saturation  salt  water  is 
measured.  If  the  boundary  is  not  at  29.6%,  slowly  turn  the  adjusting  screw  clockwise  or 
counterclockwise with the flat-bladed screwdriver supplied until the intersection is at 29.6% 
(Figure 2). 

 
(4)    After completing the adjustment, thoroughly clean the prism and the cover plate with the 

cloth supplied. 

 
(5)    Measurement 

To take a reading, place a few drops of the liquid to be measured on the prism. Look through 
the eyepiece at the white and blue fields on the scale. The concentration of the liquid being 
measured is determined by the intersection of the boundary of those fields. For example, in 
Figure 3 the concentration is 38%. 

 
 

 
Figure 1 

Figure 2 

Figure 3 

 
 
 
 
 
 

 
 
 
 
 
 
 

light

Summary of Contents for SK-100R

Page 1: ...Instruction Manual for Refractometers SK 100R SK 101R SK 102R SK 104R SK 106R SK 107R SK 109R SK 200R 3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan https www sksato co jp en H28 2...

Page 2: ...ns of this instrument When measuring substances that may be hazardous to humans make sure you understand the nature of them and wear protective gloves and mask Take special care when placing a sample...

Page 3: ...he eyepiece to see the white and blue fields as shown in Figure 2 below 3 Scale adjustment This instrument is designed to allow for scale adjustment at 20 C and the intersection of the boundary of the...

Page 4: ...iece to see the white and blue fields as shown in Figure 2 below 3 Scale adjustment This instrument is designed to allow for adjustment at 20 C and the intersection of the boundary of the white and bl...

Page 5: ...ugh the eyepiece to see the white and blue fields as shown in Figure 2 below 3 Scale adjustment This instrument is designed to allow for scale adjustment at 20 C and the intersection of the boundary o...

Page 6: ...spread the whole surface Then close the cover plate Look through the eyepiece to see the white and blue fields as shown in Figure 2 below 3 Scale adjustment This instrument is designed to allow for a...

Page 7: ...erefore the result will be 25 3 25 0 0 3 25 3 Temperature Correction Table JIS B7185 NOTES After measurement clean the prism and the cover plate to completely remove the sample otherwise it may corrod...

Page 8: ...82 0 0 5 29mm 145mm 100g 0189 00 SK 109R 0 0 to 50 0 0 5 29mm 155mm 100g 0190 00 SK 200R 0 0 to 28 0 0 2 29mm 170mm 105g for concentration of salt water Materials Main unit Aluminum alloy Cover plate...

Reviews: