background image

 

 

 

For the models SK-200R (for concentration of salt water) 
(w/ auto temperature correction) 
 

Accessories: Wipe-out cloth, carrying case, screwdriver, dropper, distilled water 

 
- Using the Refractometer after Adjustment with Distilled Water - 

 

Note: Be sure to perform the scale adjustment before using this instrument for the first time 

each day, or if the ambient temperature changes. 

 

(1)    Place yourself facing the light source and look through the eyepiece to see the scale. Adjust 

the focus to suit your eyes by rotating the eyepiece clockwise or counterclockwise. 

 
(2)    While holding the instrument as shown in Figure 1, open the cover plate, place a few drops 

of distilled water on the prism and then close the cover plate.    Look through the eyepiece to 
see the white and blue fields as shown in Figure 2 below 

 
(3)    Scale adjustment 

This instrument is designed to allow for scale adjustment at 20°C, and the intersection of the 
boundary of the white and blue fields is set to 0% when the distilled water is measured. If the 
boundary is not at 0%, slowly turn the adjusting screw clockwise or counterclockwise with 
the flat-bladed screwdriver supplied until the intersection is at 0% (Figure 2). 

 
(4)    After completing the adjustment, thoroughly clean the prism and the cover plate with the 

cloth supplied. 

 
(5)    Measurement 

To take a reading, place a few drops of the liquid to be measured on the prism. Look through 
the eyepiece at the white and blue fields on the scale. The concentration of the liquid being 
measured is determined by the intersection of the boundary of those fields. For example, in 
Figure 3 the concentration is 16%. 
 
 

 

Figure 1 

Figure 2 

Figure 3 

 
 
 
 
 
 
 
 

 
 
 

light

Summary of Contents for SK-100R

Page 1: ...Instruction Manual for Refractometers SK 100R SK 101R SK 102R SK 104R SK 106R SK 107R SK 109R SK 200R 3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan https www sksato co jp en H28 2...

Page 2: ...ns of this instrument When measuring substances that may be hazardous to humans make sure you understand the nature of them and wear protective gloves and mask Take special care when placing a sample...

Page 3: ...he eyepiece to see the white and blue fields as shown in Figure 2 below 3 Scale adjustment This instrument is designed to allow for scale adjustment at 20 C and the intersection of the boundary of the...

Page 4: ...iece to see the white and blue fields as shown in Figure 2 below 3 Scale adjustment This instrument is designed to allow for adjustment at 20 C and the intersection of the boundary of the white and bl...

Page 5: ...ugh the eyepiece to see the white and blue fields as shown in Figure 2 below 3 Scale adjustment This instrument is designed to allow for scale adjustment at 20 C and the intersection of the boundary o...

Page 6: ...spread the whole surface Then close the cover plate Look through the eyepiece to see the white and blue fields as shown in Figure 2 below 3 Scale adjustment This instrument is designed to allow for a...

Page 7: ...erefore the result will be 25 3 25 0 0 3 25 3 Temperature Correction Table JIS B7185 NOTES After measurement clean the prism and the cover plate to completely remove the sample otherwise it may corrod...

Page 8: ...82 0 0 5 29mm 145mm 100g 0189 00 SK 109R 0 0 to 50 0 0 5 29mm 155mm 100g 0190 00 SK 200R 0 0 to 28 0 0 2 29mm 170mm 105g for concentration of salt water Materials Main unit Aluminum alloy Cover plate...

Reviews: