- 20 -
⑤
Touch [Record type] and specify the type of data to be recorded.
For [Channel], dry-bulb temperature is set
CH01
and wet-bulb
temperature
is set
CH02.
・
Instant value : Records data at recording intervals.
・
Average
: Records the average value of all values measured during each recording interval.
・
Max/Min
: Records the maximum and minimum values of all values measured during each
recording interval.
Example : If the recording interval is one second, the maximum and minimum
values of 10 values measured at intervals of 100 ms will be recorded.
・
OFF
: Trend data only is recorded on the [Trend] screen and is not recorded in the
SD memory card.
Caution
・
The [Input filter], [Offset] and [Gain] settings affect measurement accuracy.
We recommend that these settings not be changed. If any of these settings are changed, the
accuracy assurance will no longer be guaranteed
.
⑥
Likewise, use [REC/CALC (Calc)] to set the recording type for relative humidity and dew point
temperature.
CH13
is set for relative humidity measurement, and
CH14
is set dew point
measurement
.