9
3 Power Supply
3.
WARNING
Turn the slide switch OFF when not in use.
3.1 External Power Supply
The PNA295-1 external power supply is 100 to 240 VAC, 45 to 65 Hz, 15 VA.
3.2 Internal Battery Power Supply
The PNA295-1 includes a rechargeable battery (12V/1.2Ah DC battery,
Yuasa NP1.2-12). When fully charged, the battery allows the PNA295-1 to
work for up to one hour. Do not allow the battery to totally discharge.
Battery Storage
Long-term storage of the battery is according to the following:
Storage temperature
+10
°
C +20
°
C +30
°
C +40
°
C
Time before recharging required,
in months
>
18
14 8 5
Charging the Battery
When the battery is fully discharged (indicated by the blinking LOW BAT
lamp or by no display), apply AC voltage to the PNA295-1 to recharge the
battery. The recommended charge time is 8 hours, which will allow at least
20 minutes of work.
ATTENTION
For reconnection to the AC network and charging, turn the instrument slide switch
OFF, wait approximately 5 minutes and then turn it ON again.
Summary of Contents for PNA295-1
Page 1: ...PNA295 1 Portable Network Analyzer Installation and Operation Manual BG0149 Rev B...
Page 2: ...PNA295 1 Portable Network Analyzer Installation and Operation Manual...
Page 18: ...13 Figure 2 3 PNA295 1 Connections for 4 Wire System Direct Connection...
Page 19: ...14 Figure 2 4 PNA295 1 Connections for 4 Wire Grounded Delta System Direct Connection...