Satec PM295 Installation And Operation Manual Download Page 12

4

Introduction

16 programmable data logging partitions, each for recording from 1 to
16 user programmable parameters per record, providing total storage
for up to 114,688 parameters with a 512K memory module (assuming
the entire memory allocated for data logging)

one partition for high-speed waveform recording (32 samples x 16
cycles x 6 inputs per record) for disturbance analysis, and one partition
for high-resolution waveform recording (128 samples x 4 cycles x 6
inputs per record) for harmonic analysis, each providing storage for up
to 19, 40, or 82 records with a 512K memory module (assuming the
entire memory allocated for the partition)

 

each memory partition can be sized to store desired number of records, from

1 record and up to the entire memory size, using an arbitrary combination of
partitions

 

each memory partition can be programmed to store the oldest records without
overwriting previously recorded data when a partition is filled up, or to wrap
around by writing new records over the oldest records so that stored records
will always contain the most recent data

Time-of-Use System

 

8 programmable accumulating energy registers for 16 tariffs, each
configurable for counting kWh/kvarh/kVAh or pulses from up to 8 external
energy-counting meters

 

3 programmable demand registers for 16 tariffs, configurable for recording
extreme (minimum and maximum) demands using varying calculation
techniques: block interval/sliding window/thermal demand

 

16 daily profiles (types of days) with up to 8 tariff changes per day

 

2-year calendar

Communications

 

one programmable optically isolated serial port - RS-232/RS-422/RS-485
embedded options with a user-selectable baud rate of 110 to 38400 BPS

Display

 

multi-page display composed of 11 windows with high-brightness seven-
segment digital LEDs. A total of 55 display pages are available

Keypad

 

four membrane long-life push-buttons for page scrolling and programming

Power Supply

 

90-264 V AC,  10-290 V DC options

Summary of Contents for PM295

Page 1: ...System 295 Powermeter Harmonic Analyzer Installation and Operation Manual BG0130 Rev A2...

Page 2: ...SYSTEM 295 POWERMETER HARMONIC ANALYZER Installation Operation Manual...

Page 3: ...ver these instructions do not cover all possible contingencies that may arise during installation operation or maintenance and all details and variations of this equipment are not covered by these ins...

Page 4: ...duce mechanical strain on the screw terminals if necessary 6 Setup procedures must be performed only by qualified personnel familiar with the instrument and its associated electrical equipment 7 DO NO...

Page 5: ...e and shows the corresponding front panel displays Appendix B lists the instrument programmable parameters that can be accessed either through the front panel or communications Appendix C provides a s...

Page 6: ...s 26 2 11 Relay Output Connections 27 2 12 Discrete Input Connections 28 2 13 Communications 28 3 Operating The PM295 30 3 1 Instrument Turn On 30 3 2 Operational Mode 30 3 2 1 Front Panel Operation 3...

Page 7: ...ents 63 4 4 1 Measurement Modes 63 4 4 2 Resetting the Energies 64 4 4 3 Energy Pulsing 64 4 5 Harmonic Measurements 64 4 5 1 Measurement Technique 64 4 5 2 Harmonic Parameters 65 4 5 3 Real time Wave...

Page 8: ...18 Update Rates And Response Time 99 5 Communications Operation 101 5 1 General 101 5 2 Eia Interface Standards 101 5 2 1 EIA RS 232 Standard 101 5 2 2 EIA RS 422 and EIA RS 485 Standards 102 5 3 Conf...

Page 9: ...directly by transducing analog and discrete signals serving as excellent partner in power network management All measurement functions of the PM295 are remotely programmable via the communications po...

Page 10: ...g parameters that can be controlled externally by a combination of binary signals on the instrument s discrete inputs from 1 to 4 inputs can be operated extension for up to 14 analog outputs is availa...

Page 11: ...mulating energy registers reset TOU system demand registers clear all counters clear Min Max log registers event logging on setpoint operation data logging in selected data log partition high speed wa...

Page 12: ...t overwriting previously recorded data when a partition is filled up or to wrap around by writing new records over the oldest records so that stored records will always contain the most recent data Ti...

Page 13: ...rement parameters are summarized in technical specifications see Chapter 6 Chapter 3 describes measurement functions and modes of measuring input values Table 1 1 Measured Quantities and Sensed Signal...

Page 14: ...nt Total power factor lead Auxiliary measurements Auxiliary current ground leakage neutral current Neutral current calculated Frequency Voltage unbalance Current unbalance Low values on any phase Low...

Page 15: ...r lead High voltage THD High current THD High K Factor Demands Volt demand per phase Ampere demand per phase Total kW demand block interval sliding window thermal Total kvar demand block interval slid...

Page 16: ...oltages for odd harmonics 1 39 Harmonic currents for odd harmonics 1 39 Three phase total harmonic measurements Harmonic total kW for odd harmonics 1 39 Harmonic total kvar for odd harmonics 1 39 Harm...

Page 17: ...large scale counters for counting external pulses or internal events Discrete inputs 8 discrete inputs free configurable for sensing external contacts or pulses Timers 4 timers with 1 sec resolution I...

Page 18: ...r will be zero because they have no meaning Only total three phase power values can be used In all grounded connections using either 4Ln3 or 4LL3 wiring configuration s harmonic voltages will represen...

Page 19: ...It must be placed at least one half meter 1 64 feet from current lines carrying up to 600 amperes For currents greater than 600A and up to 2 000A this distance must be at least 1 meter 3 28 feet In th...

Page 20: ...12 Installation Figure 2 1 Front Mounting standard...

Page 21: ...Installation 13 Figure 2 2 Rear Mounting Step 1 Connect the bracket to the instrument Step 2 Mount the O ring on the instrument Step 3...

Page 22: ...als Connections to the PM295 are made via terminals located on the back of the instrument as shown inFigure 2 3 and detailed inFigure 2 4 Figure 2 3 Location of Terminal Strips and Communications Conn...

Page 23: ...pend on the power system configuration and on the system voltage level In the case of systems with line to line voltage up to 660V direct connection may be used for 4 wire and 3 wire systems Wiring di...

Page 24: ...galvanically isolated using internal current transformers Using 3 wire connections only two current transformers are required It is also possible to use three current transformers seeFigure 2 7 If nec...

Page 25: ...be calculated inaccurately 3 wire Direct Connection In a 3 wire Direct Connection harmonicvoltages will represent the three phaseline to neutral voltages that appear on the instrument input transform...

Page 26: ...nnection using 2 PTs 3 CTs 2 element 3OP3 4 4 wire WYE direct connection using 3 CTs 3 element 4Ln3 or 4LL3 5 4 wire delta direct connection using 3 CTs 3 element 4Ln3 or 4LL3 6 4 wire WYE connection...

Page 27: ...voltage up to 660V The three line voltages are taken at terminals 2 5 and 8 as shown inFigure 2 5 Figure 2 5 3 wire Direct Connection Using 2 CTs 2 element WIRING MODE 3DIR Readings represent line to...

Page 28: ...Readings represent line to line voltages The two line currents are measured via 2 CTs the third current LINE 2 B is calculated based on the two measured currents The common taps of the PTsecondaries a...

Page 29: ...can be used with either 660V or 120V input Readings representline to line voltages All three line currents are measured The common taps of the PTsecondaries are connected to terminal 11 Note the conne...

Page 30: ...ing 3 CTs 3 element The instrument takes the three line to neutral voltages and three line currents as shown inFigure 2 8 The system neutral is connected to terminal 11 Figure 2 8 4 Wire Wye Direct Co...

Page 31: ...ng 3 CTs 3 element The instrument senses the three line to neutral voltages and three line currents as shown inFigure 2 9 The system neutral is connected to terminal 11 Figure 2 9 4 wire Delta Direct...

Page 32: ...be used with either 660V or 120V input The instrument senses the three line to neutral voltages and three line currents as shown in Figure 2 10 The common taps of the PTsecondaries are connected to te...

Page 33: ...e 2 11 4 wire Wye Connection Using 2 PTs 3 CTs 2 element WIRING MODE 3Ln3 3LL3 2 9 Auxiliary Current Input Connections The PM295 can be equipped with an auxiliary current input optional Figure 2 12 sh...

Page 34: ...an internal source 24 VDC to powerthe current loop Analog output range is 0 20 or 4 20 mA as per your order R L max 510 Ohm 15 13 Figure 2 13 Analog Output Connections The permitted range for the curr...

Page 35: ...e relays All relays are normally de energized 19 23 K3 21 K1 K2 22 24 26 28 K4 25 27 Figure 2 15 Relay Output Connections Relay Contact Protection When using relay outputs for switching lamp capacitiv...

Page 36: ...s isolated from noise sources 2 13 Communications Connector Pinout The communications port is optically isolated and supports both EIA RS 232and RS 422 485 standard interfaces user selectable The seri...

Page 37: ...nsmit line and flashes when the instrument is transmitting data Cable Connections For the cable drawings refer toAppendices D E and F For RS 422 RS 485 balanced data transmission a shielded twisted pa...

Page 38: ...ferent diagnostic code continually appears when you apply power to the instrument contact your local distributor For error codes refer to Section 3 4 Upon power up the PM295 assumes the operational mo...

Page 39: ...ctive page number is indicated on the display It illuminates continuously when you are on the main level and flashes when you enter the sub page level At power up the instrument always returns to the...

Page 40: ...ted level can view modify setups and perform management functions reset RTC update superuser level privileged level all functions allowed at the protected level plus user protection override This leve...

Page 41: ...ot be accessed with the front panel keys button window a labeled window allowing the user to perform an action to enter or quit a menu Button window acts as labeled button and displays an abbreviated...

Page 42: ...ard To enter the selected value When desired item or value appears in the window press ENTER If the value in the currently active window is selected correctly after pressing ENTER you will exit the wi...

Page 43: ...ted level If password protection is enabled you will be prompted to enter a password 3 3 4 Entering Quitting Programming Mode To enter programming mode from operational mode Press SELECT The first men...

Page 44: ...RETE INPUT SETUP Cnt COUNTER SETUP Aout AEPn ANALOG OUTPUT SETUP PulS SEtP ALARM EVENT SETPOINTS t r TIMER SETUP rtc RTC SETUP diSP rSt RESET CLEAR AccS PROTECTION CONTROL SELECT PASS SEE CHG ESC OOOO...

Page 45: ...t with the up down arrow keys In the same manner set the other password digits Press ENTER 3 3 6 Selecting the Setup Group The setup parameters are organized into 13 groups accessed via primary menus...

Page 46: ...l the label of the desired parameter appears The parameter value will be indicated in the lower window When the parameter value exceeds the number of places in the window the high order digits are exp...

Page 47: ...meters with the up down arrow keys until the label of the desired parameter appears The parameter value will be indicated in the lower window To change the setup parameter Press SELECT to choose the l...

Page 48: ...ENTER to return to the upper window To quit the menu From the upper window pressENTER to return to theMAIN menu The illustrations above show an example of discrete inputsallocation Here inputs 1 5 are...

Page 49: ...keys The nonE entry disables external input To change the scale factor for the counter Press SELECT to choose the lower window Adjust the scale factor for the counter with the up down arrow keys The...

Page 50: ...able B 4a see Appendix B with the corresponding group label The analog output parameter is defined by both the parameter group and parameter name within the group All applicable parameters are listed...

Page 51: ...u TROUBLESHOOTING 1 If your setup for either scale isn t accepted by the instrument check whether the low scale value does not exceed the parameter full scale 2 The output scales for the signed power...

Page 52: ...select a relay output setup From the upper window select the desired relay with the up down arrow keys To change the output parameter for the relay Press SELECT to choose the central window Scroll thr...

Page 53: ...disable the setpoint or to replace the old setup with the new configuration The lower window allows to exit the menu and return to theMAIN menu ESC nonE SP 1 To select a setpoint setup From the upper...

Page 54: ...e lower window To quit the menu Press SELECT to choose the ESC window Press ENTER to return to theMAIN menu Viewing and Changing the Setpoint Configuration To view or change the setpoint configuration...

Page 55: ...one of the entriesCnd1 to Cnd4 in the upper window using the up down arrow keys then pressSELECT The following illustrations show the sub menu windows Cnd1 ConJ Or Cnd1 GrP rt Hi Cnd1 InP cur Cnd1 Co...

Page 56: ...ys until the desired entry appears Press ENTER to return to the central window From the central window select theInP entry with the up down arrow keys The lower window will displaythe first trigger pa...

Page 57: ...t parameter setup Act1 tYPE rEL Action type Act1 tArG rEL 1 Action target For applicable action types and their appropriate targets refer toTable B 9 see Appendix B To specify setpoint actions refer t...

Page 58: ...rrow keys The lower window displays the current parameter setup To change the delay unit From the central window select theUnit entry with the up down arrow keys Press SELECT to choose the lower windo...

Page 59: ...se the lower window Adjust the timer interval with the up down arrow keys The applicable range is 0 to 9999 sec Setting the interval to zero disables the timer run Press ENTER to return to the central...

Page 60: ...eys Press ENTER to return to the upper window To reset seconds Press SELECT to choose the second s window Press ENTER to return to the upper window Date Indication The date is displayed in the user se...

Page 61: ...n d Y To set the desired date format Press SELECT to choose the desired position Select the appropriate item for the position with the up down arrow keys Press ENTER to return to the upper window To...

Page 62: ...to the central window To quit the menu From the central window pressENTER to return to theMAIN menu 3 3 19 Password Protection Control The PASSWORD PROTECTION CONTROL menu is visible only at the prote...

Page 63: ...T to choose the lower window With the up down arrow keys select OFF to disable password protection and select On to enable password protection Press ENTER to return to the central window 3 4 Self Test...

Page 64: ...t is possible to capture data acquired prior to a trigger signal to provide pre and post event waveform analysis Sampling frequency is periodically accelerated from a general rate to an expanded rate...

Page 65: ...chniques of time averaging block interval demand sliding window demand and thermal demand For each method calculations are performed within the user definedDEMAND PERIOD The measured quantity is integ...

Page 66: ...and apparent kVA power over all three phases and volt and ampere demand s per phase All demand readings are represented by a positive number Active power demand is calculated for positive power only...

Page 67: ...measured total powers This method is similar to the way electro mechanical thermal demand meters operate The three phase thermal maximum demand measuring element comprises a thermally lagged power de...

Page 68: ...stant will be 195 4 seconds using a 15 minute billing demand interval and 390 9 seconds using a 30 minute billing demand interval 4 3 5 Accumulated and Predicted Demands The PM295 offers two evaluated...

Page 69: ...terval over which extrapolation is made The predicted demand deviation might be slightly increased near the beginning of the demand interval where a base for extrapolation is too small In the event th...

Page 70: ...the instrument s RTC readings In the event of loss of power or changing any demand parameter the instrument immediately begins a new shorter demand interval until the first synchronization For instan...

Page 71: ...r changing the demand setup and after power up the demand interval will have a maximum value allowed of one hour Until the second synchronization pulse the accumulated demand readings will be incorrec...

Page 72: ...resetcan be performed concurrently in all instruments connected to a master computer The date and time of the last resetcan be read via the instrument s front panel 4 4 3 Energy Pulsing Each of four r...

Page 73: ...p to 39th real time harmonictotal powers active and reactive for odd harmonics up to 39th real time harmonicpower factors for odd harmonics up to 39th Total harmonic distortions THD and K Factors are...

Page 74: ...1 For voltage waveforms see NOTE 1 in the previous paragraph 2 For a 3 wire connection sampled waveforms may not be used for calculation of power harmonics on separate phase Notice that high resolutio...

Page 75: ...ption can be ordered with the secondary rating for the external CT with a 5 mA 1 A or 5A full scale secondaries Ground Leakage Measurements The auxiliary input is typically used for ground leakage mea...

Page 76: ...tation order is defined using the phase A and C voltages The phase rotation order can be read via the front panel and used as a trigger for setpoint operations 4 6 6 Phase Angles To assist the user in...

Page 77: ...te date before daylight savings time becomes effective and at daylight savings time ending 4 7 2 TOU System Registers Accumulating Energy Registers Energy measurements are provided for eight energysou...

Page 78: ...utput Starting a new tariff interval is also considered to be an internal event that can trigger setpoint operation for instance in order to synchronize self readings data recording with beginning of...

Page 79: ...e made in response to external pulse Third one of the pulse inputs can be configured to sense the external synchronization pulse in the event that external time reference is used for the demandinterva...

Page 80: ...tput Options The PM295 has one free scalable current output 0 20 4 20 mA upon order that provides transmitting a current proportional to the measured quantity through external load current loop The an...

Page 81: ...hin the parameter range can be specified The only exception is the signed power factor for which the output scales are set permanently to the range of 1 00 to 1 00 and the center of the scale correspo...

Page 82: ...that has seven current outputs 0 20 4 20 mA and is connected to the Powermeter via a RS 422 communications link It can share the same link with the master computer or PLC without interfering in their...

Page 83: ...imum value allowed or reverse up to zero order Each counter input can be independently scaled weighted by specifying a scale factor in the range of 1 to 9999 units per pulse This means that each count...

Page 84: ...l control over setpoint operations and for more sophisticated setpoint programming Each event can be asserted or cleared independently by setting ON or clearing OFF the dedicated event flag The user c...

Page 85: ...ints he must hold the timer event until the last setpoint is checked Using a programmable event flag the user can store the volatile event by setting the flag by a dedicated setpoint and then continuo...

Page 86: ...each partition in the memory module does not depend on its order in a list of partitions Each time the user reconfigures the extended memory the instrument performs memory optimization to keep free me...

Page 87: ...gs into the extended memory are programmed by the user and can be made only through the setpoint operation The only exception is the case when the instrument detects a situation that affects its opera...

Page 88: ...on the event occurrence event target identifying the point to which the event action is intended 4 15 2 Data Logging The PM295 provides data logging in 16 independent memory partitions each programmab...

Page 89: ...r response and the instrument may not guarantee that the event that caused the trigger to operate will exist within the waveform buffer at the time that waveform recorder runs In this case recorded wa...

Page 90: ...ycle The voltage and current waveforms on each separate phase are always recorded synchronously and can be used for power harmonic measurements Waveforms corresponding to different phases are not sync...

Page 91: ...allowed in percentage of nominal full scale voltage The reference nominal voltage is 120V RMS 170V amplitude for instruments with the 120V input option and 380V RMS 537V amplitude for instruments wit...

Page 92: ...led to make a decision concerning the setpoint operation The event processor can provide a one shot action on operating the setpoint as well as periodical actions using one of the four interval timers...

Page 93: ...liable operation on alarm conditions The following sections provide a more detailed explanation of the setpoint programming technique and setpoint operations 4 17 3 Triggering Conditions Each setpoint...

Page 94: ...d band on event operation When the trigger parameter value is between the operate and release limits the event statusdoes not change If the event was asserted before entering the dead band it is still...

Page 95: ...itions evaluated before it when bothOR and AND operations are combined in one expression As example any trigger condition bound with anOR operation and being evaluated to true will override any preced...

Page 96: ...se the release conditions will never be realized so that the setpoint will be locked in the operate state permanently To release such a setpoint you will need to set it up again 2 When you are using a...

Page 97: ...on Data logging Log number 1 16 High speed 32 16 waveform logging n a 0 High resolution 128 4 waveform logging n a 0 All actions are always run in the order they are enumerated in the setpoint specifi...

Page 98: ...s operation will be different from that was prior to loss of power Programming Mode When the user enters programming mode at the protected level allowing the setup to be changed setpoint monitoring is...

Page 99: ...eds its predefined operate limit the setpoint is triggered and a 3 second delay timer runs If the measured value remains away from the threshold for minimum of 3 seconds the setpoint is operated Three...

Page 100: ...d parameters in data log partition 2 the set of parameters should be specified for the partition separately on each new maximum block kW demandoccurrence while status input 5 remains asserted No delay...

Page 101: ...we are intending to provide resetof extreme demands on the first day of each month synchronously with an external time reference pulse denoting beginning of the new demand interval The demand interval...

Page 102: ...ollowing setpoint will then provide all needed operations If a data logging partition is configured to be of a non wrap type recording will stop when the partition is filled up Otherwise logging will...

Page 103: ...5 on the 12th Aug 1997 from 08 30 00 to 17 15 00 Example 4 7 Sample Of Repeated Time Gated Operations Trigger conditions Conjunction Trigger parameter Condi tion Operateli mit Releaseli mit Condition...

Page 104: ...ed operate the setpoint by overriding other setpoint conditions Clearing the event flag does not ensure that the setpoint will be released if other operate conditions are still fulfilled Example 4 9 S...

Page 105: ...either phase over this limit will trigger high speed recording the captured waveforms in the corresponding memory partition The operate limit is selected large enough to prevent occasional setpoint op...

Page 106: ...n it comes in the event flag if the event has not yet occurred Two setpoints 2 and 3 are triggered by the event flag and provide continuous recording data using interval timer 1 Setpoint 4 disables re...

Page 107: ...ents performed at the instrument inputs Response time is the time between a change in an input and when that change is displayed via the front panel or provided on the instrument outputs analog output...

Page 108: ...for averaging For time average measurements the update rate is synchronized with the user defined averaging time or with the external synchronization source All time average measurements are updated a...

Page 109: ...y host based software to access all data and configuration registers of the Powermeter Information on the serial communications protocols is found in the documents shipped on diskette with yourPM295 S...

Page 110: ...al connector pins to which the communications line is connected Table B 3 see Appendix B lists the communications parameters that should be defined by the user These parameters must have the same valu...

Page 111: ...ter units connected to a single communications line each Powermeter on the network must have its own unique address Choices for address depend on communications protocol you are using In ASCII communi...

Page 112: ...tware handshaking When hardware handshaking is selected the Powermeter will not send characters to the communications link until the DSR CTS signal is asserted high Whenever DSR CTS dropped while tran...

Page 113: ...d as those on the printer Cable connections to the printer may depend on the printer connector type Most of serial printers have a DTE 25 pin male connector but some of them might have a DCE 25 pin fe...

Page 114: ...s link plus the link switching time The communications link switching time is about 15 ms plus 1 75 character time The transmission of the analog output message for one AX 7 s channel takes 9 characte...

Page 115: ...he print report are listed in Table 5 1 Table 5 1 Printout Format Field number Parameter Description Offset in the string 1 2 3 V1 V2 V3 Voltage per phase 9 17 25 4 5 6 A1 A2 A3 Current per phase 33 4...

Page 116: ...RMS for 1 second Auxiliary current input 5 mA INPUT VIA CT with 5 mA secondary output Burden 0 1 mVA Overload withstand 30 mA RMS continuous 400 mA RMS for 1 second 1 A INPUT via CT with 1 A secondary...

Page 117: ...t terminals Standard 9 52 mm pitch UL recognized 7463 SCREW M3 Maximum wire diameter 2 588 mm 10 AWG Service terminals Standard 5 mm pitch UL recognized 7463 SCREW M3 Maximum wire diameter 2 05 mm 12...

Page 118: ...100 to 2 147 MW 0 1 10 to 2 147 MW 1 kW Reactive power kvar 0 36 PT CT 120V input 1 14 PT CT 660V input 0 5 PF 0 9 0 to 2 147 000 1 kvar 1 to 99 999 9 999 0 01 100 to 2 147Mvar 0 1 10 to 2 147 Mvar 1...

Page 119: ...demand V As for voltage Ampere demand A As for current kW demand kW As for kW kvar demand kvar As for kvar kVA demand kVA As for kVA Active energy kWh 1 typical 0 to 999 999 999 1 kWh 1 kWh Reactive e...

Page 120: ...al As for power factor PT the external potential transformer ratio CT CT primary current the primary current rating of the external current transformer FSV voltage full scale FSI current full scale U1...

Page 121: ...A TOTAL 1 4 10 5 6 7 8 9 3 2 11 1 V L1 L12 CURRENT L1 kVA total L1 PF L1 kW L1 kvar L1 kVA L1 1 4 10 5 6 7 8 9 3 2 11 1 V L2 L23 CURRENT L2 kVA total L2 PF L2 kW L2 kvar L2 kVA L2 1 4 10 5 6 7 8 9 3 2...

Page 122: ...D DEMAND 1 4 10 5 6 7 8 9 3 2 11 kVA total 2 MAX kVA DEMAND MAX kW MAX kvar DEMAND DEMAND tr d 1 4 10 5 6 7 8 9 3 2 11 kVA total 2 Pr d kVA DEMAND kW DEMAND kvar DEMAND 1 4 10 5 6 7 8 9 3 2 11 kVA tot...

Page 123: ...4 10 5 6 7 8 9 3 2 11 kVA total k W h 3 EnrG Ac i I M P O R T 1 4 10 5 6 7 8 9 3 2 11 kVA total k W h 3 EnrG Ac E E X P O R T 1 4 10 5 6 7 8 9 3 2 11 kVA total k W h 3 EnrG Ac n N E T Page 3 THD neutr...

Page 124: ...rE i I M P O R T 1 4 10 5 6 7 8 9 3 2 11 kVA total 3 EnrG k v a r h rE E E X P O R T 1 4 10 5 6 7 8 9 3 2 11 kVA total 3 EnrG k v a r h rE n N E T Page 3 4 kWh total extended range windows 3 6 9 Page...

Page 125: ...T A L 1 4 10 5 6 7 8 9 3 2 11 kVA total 3 EnrG Ap t k V A h T O T A L 1 4 10 5 6 7 8 9 3 2 11 kVA total 3 EnrG date reset D A T E T I M E Page 3 8 kvarh total extended range windows 3 6 9 Page 3 9 kV...

Page 126: ...kVA total V L3 H03 H39 4 PF TOTAL kW TOTAL kvar TOTAL 1 4 10 5 6 7 8 9 3 2 11 kVA total 5 K FACTOR L1 K FACTOR L2 K FACTOR L3 H Fc 1 4 10 5 6 7 8 9 3 2 11 kVA total 5 Ph ro U Unb C Unb PHASE ROTATION...

Page 127: ...8 9 3 2 11 kVA total 5 rEL RELAY 4 STATUS RELAY 1 RELAY 2 RELAY 3 STATUS STATUS STATUS 1 4 10 5 6 7 8 9 3 2 11 kVA total 5 Ph AG PHASE L1 PHASE L2 PHASE L3 ANGLE ANGLE ANGLE dEG 1 4 10 5 6 7 8 9 3 2...

Page 128: ...1 7 3 Power factor L1 1 1 8 5 kW L1 kW MW 1 1 9 5 kvar L1 kvar Mvar 1 1 10 4 kVA L1 kVA MVA 1 2 2 3 Voltage L2 L23 V kV 1 2 5 4 Current L2 A kA 1 2 7 3 Power factor L2 1 2 8 5 kW L2 kW MW 1 2 9 5 kva...

Page 129: ...g win kvar demand kvar Mvar 2 3 10 4 Predicted sliding win kVA demand kVA MVA 2 4 8 5 Accumulated kW demand kW MW 2 4 9 5 Accumulated kvar demand kvar Mvar 2 4 10 4 Accumulated kVA demand kVA MVA 3 1...

Page 130: ...total kW kW MW 4 9 5 H01 harmonic total kvar kvar Mvar 4 1 19 1 3 H03 H39 harmonic voltage L1 L12 V kV 4 1 19 2 3 H03 H39 harmonic voltage L2 L23 V kV 4 1 19 3 3 H03 H39 harmonic voltage L3 V kV 4 1 1...

Page 131: ...icparameters pages 2 and 4 are sliding average values Voltage and current readings with a decimal point are displayed in kV kA Auxiliary current readings with a decimal point are displayed in amperes...

Page 132: ...to 1111111 Counter setup Counter number Selects one of the eight pulse counters 1 to 8 Discrete input number Connects one of the eight discrete pulse inputs to the counter 1 to 8 Scale factor Defines...

Page 133: ...ter Specifies pulsing parameter see Table B 5 Unit hours per pulse Defines the number of unit hours per pulse for energy pulses 0 1 to 9999 Event alarm setpoints Setpoint number Selects one of the six...

Page 134: ...he time delay to release the setpoint 1 to 9999 Timer setup Timer number Selects one of the four interval timers 1 to 4 Timer interval Defines the time interval for the timer 1 to 9999 sec Real time c...

Page 135: ...log setup Register number Selects the programmable Min max log register to be setup 1 to 16 Measurement parameter group Specifies the parameter group for the register Measurement parameter Specifies t...

Page 136: ...ly profiles types of days 1 to 16 Tariff change point Selects one of the eight tariff change points Tariff change time Specifies the tariff start time Tariff number Selects one of the sixteen tariffs...

Page 137: ...rrent 1 50000 mA A The primary rating of the auxiliary current transformer d P Demand period 1 2 5 10 15 20 30 60 E The length of the demand period for power demand calculations minutes E external syn...

Page 138: ...110 baud 300 300 baud 600 600 baud 1200 1200 baud 2400 2400 baud 4800 4800 baud 9600 9600 baud 19 2 19 200 baud 38 4 38 400 baud dAtA Data format 7E 7 bits even parity 8n 8 bits no parity 8E 8 bits e...

Page 139: ...oltages odd harmonics hr C1 L1 phase harmonic currents odd harmonics hr C2 L2 phase harmonic currents odd harmonics hr C3 L3 phase harmonic currents odd harmonics hr Ac Harmonic total active powers od...

Page 140: ...Voltage THD L3 0 100 0 th C1 Current THD L1 0 100 0 th C2 Current THD L2 0 100 0 th C3 Current THD L3 0 100 0 HFc 1 K Factor L1 n a 1 0 999 9 HFc 2 K Factor L2 n a 1 0 999 9 HFc 3 K Factor L3 n a 1 0...

Page 141: ...AP P1 kVA L1 kVA 0 Pmax AP P2 kVA L2 kVA 0 Pmax AP P3 kVA L3 kVA 0 Pmax PF1 Power factor L1 n a 1 00 1 00 PF2 Power factor L2 n a 1 00 1 00 PF3 Power factor L3 n a 1 00 1 00 th U1 Voltage THD L1 L12...

Page 142: ...Ad 3 Ampere demand L3 A 0 Imax Ac bd Block kW demand import kW 0 Pmax rE bd Block kvar demand total kvar 0 Pmax AP bd Block kVA demand kVA 0 Pmax Ac Sd Sliding window kW demand import kW 0 Pmax rE Sd...

Page 143: ...0 Imax hr Ac Harmonic total active powers odd harmonics H01 H39 Harmonic H01 H39 kW Pmax Pmax hr rE Harmonic total reactive powers odd harmonics H01 H39 Harmonic H01 H39 kvar Pmax Pmax hr PF Harmonic...

Page 144: ...C Special inputs uSr E User event flags manual control int E Internal events t r Timers St In Status inputs PS In Pulse inputs rEL Relay status Cnt Counters ti E Time date parameters rt Ph Real time p...

Page 145: ...ts odd harmonics hr Ac Harmonic total active powers odd harmonics hr rE Harmonic total reactive powers odd harmonics hr PF Harmonic total power factors odd harmonics Lo Ph New minimum real time values...

Page 146: ...events On OFF Ac Ei kWh import pulse Ac EE kWh export pulse Ac Et kWh total pulse rE Ei kvarh import pulse rE EE kvarh export pulse rE Et kvarh total pulse AP Et kVAh total pulse dint Demand interval...

Page 147: ...1 V 0 to Vmax cur1 Current L1 A 0 to Imax cur2 Current L2 A 0 to Imax cur3 Current L3 A 0 to Imax Ac P1 kW L1 kW Pmax to Pmax Ac P2 kW L2 kW Pmax to Pmax Ac P3 kW L3 kW Pmax to Pmax rE P1 kvar L1 kvar...

Page 148: ...Low voltage THD 0 to 100 0 thd C Low current THD 0 to 100 0 HFc Low K Factor 1 0 to 999 9 rt Hi Real time high values on any phase GE LE Eq nE U High voltage V 0 to Vmax cur High current A 0 to Imax...

Page 149: ...0 to Imax cur3 Current L3 A 0 to Imax Ac P1 kW L1 kW Pmax to Pmax Ac P2 kW L2 kW Pmax to Pmax Ac P3 kW L3 kW Pmax to Pmax rE P1 kvar L1 kvar Pmax to Pmax rE P2 kvar L2 kvar Pmax to Pmax rE P3 kvar L3...

Page 150: ...ues on any phase GE LE Eq nE U High voltage V 0 to Vmax cur High current A 0 to Imax Ac P High kW kW Pmax to Pmax rE P High kvar kvar Pmax to Pmax AP P High kVA kVA 0 to Pmax PF LG High power factor l...

Page 151: ...Sliding window kvar demand kvar 0 to Pmax AP Sd Sliding window kVA demand kVA 0 to Pmax Ac td Thermal kW demand kW 0 to Pmax rE td Thermal kvar demand kvar 0 to Pmax AP td Thermal kVA demand kVA 0 to...

Page 152: ...to Vmax hr C1 L1 phase harmonic currents odd harmonics GE LE Eq nE H01 H39 Harmonic H01 H39 A 0 to Imax hr C2 L2 phase harmonic currents odd harmonics GE LE Eq nE H01 H39 Harmonic H01 H39 A 0 to Imax...

Page 153: ...L23 th U3 Voltage THD L3 th C1 Current THD L1 th C2 Current THD L2 th C3 Current THD L3 HFc 1 K Factor L1 HFc 2 K Factor L2 HFc 3 K Factor L3 Lo to New minimum real time total values neU Ac P Total kW...

Page 154: ...g window kvar demand AP Sd Sliding window kVA demand Ac td Thermal kW demand rE td Thermal kvar demand AP td Thermal kVA demand Lo PG Programmable Min Max minimum registers nEU rG 01 rG 16 Programmabl...

Page 155: ...Factor L3 Hi to New maximum real time total values neU Ac P Total kW rE P Total kvar AP P Total kVA PF Total PF PF LG Total PF Lag PF Ld Total PF Lead Hi Au New maximum real time auxiliary values nEU...

Page 156: ...tF01 tF16 ProF Active profile Pr01 Pr16 Lt Ac New TOU System minimum kW demands nEU tF01 tF16 Tariff 1 16 Lt rE New TOU System minimum kvar demands nEU tF01 tF16 Tariff 1 16 Lt AP New TOU System mini...

Page 157: ...120V input option and 380V RMS for instruments with the 660V input option Table B 8 Setpoint Conditions Label Operate condition Release condition Limits GE Greater or equal over operate limit Less or...

Page 158: ...logging dt LG Data logging LG01 LG16 Log number LF LG High speed 32 16 waveform logging nonE n a HF LG High resolution 128 4 waveform logging nonE n a Table B 10 Reset clear Functions Access Function...

Page 159: ...lution 128 4 waveform log Clears the high resolution waveform log partition Restore event log Restores the event log queue Restore data log Restores the read queue for one of the 16 data log partition...

Page 160: ...ndition 1 OR Condition 2 Condition 3 Condition 4 Setpoint actions Delays Action type Action target Unit Operate Release Action 1 Action 2 Action 3 Action 4 Setpoint Trigger conditions Conjunction Trig...

Page 161: ...he PC cable end make a short between DSR DTR pins and between RTS CTS pins 25 pin Computer Connector 9 pin DB9 male connector 25 pin DB25 female connector 1 2 3 Powermeter TxD RxD RxD TxD 7 3 2 6 5 5...

Page 162: ...C cable end make a short between RTS CTS pins 25 pin Computer Connector 9 pin DB9 male connector 25 pin DB25 female connector 1 2 3 Powermeter TxD RxD RxD TxD 7 3 2 6 5 5 4 20 DSR CTS GND GND IBM PC C...

Page 163: ...hain and on the master computer 2 The shield of each segment of the cable must be connected to the ground at one end only 3 Avoid star or stub connected topologies 25 pin Computer Connector RxD TxD Rx...

Page 164: ...ion 9 pin Computer Connector RxD TxD RxD 120 32 TxD Shield RxD TxD RxD TxD RxD 2 TxD RxD 1 TxD TxD RxD RTS CTS CTS RTS Shield 5 4 RxD TxD Shield 120 3 Powermeters 8 9 2 7 6 A 9 pin DB9 female connecto...

Page 165: ...hain and on the master computer 2 The shield of each segment of the cable must be connected to the ground at one end only 3 Avoid star or stub connected topologies 25 pin Computer Connector RxD RxD Tx...

Page 166: ...ion 9 pin Computer Connector RxD RxD TxD 32 TxD 120 Shield TxD RxD RxD TxD RxD 2 RxD TxD Shield 1 TxD Shield 2 CTS RxD TxD CTS RTS RTS 120 5 4 TxD RxD 3 IBM PC compatible Powermeters A 9 pin DB9 femal...

Page 167: ...s than 256 bytes of input buffer At the printer s cable end make a short between DSR and DTR pins 25 pin DTE Printer Connector 1 2 3 Powermeter Printer TxD RxD RxD TxD 7 3 2 6 5 5 4 20 DSR DTR CTS GND...

Page 168: ...connector RTS 4 DSR 6 DSR CTS DTR RTS In some printers an additional short may be needed between DSR and RTS pins dashed line 9 pin DCE Printer Connector 1 2 3 Powermeter Printer TxD RxD RxD TxD 7 3...

Page 169: ...between DSR and CTS pins may be needed dashed line 25 pin DTE Printer Connector 1 2 3 Powermeter Printer TxD RxD RxD TxD 7 3 2 6 5 5 4 20 DSR DTR CTS GND GND A 9 pin DB9 male connector A 25 pin DB25 f...

Page 170: ...r TxD RxD RxD TxD 7 3 2 5 4 20 DTR GND GND A 9 pin DB9 male connector A 25 pin DB25 male connector RTS 4 DSR 6 DSR CTS DTR RTS 9 pin DCE Printer Connector 1 2 3 Powermeter Printer TxD RxD RxD TxD 7 3...

Page 171: ...control option Refer to your modem manual for appropriate commands In most cases you can use the following command string to set up your remote modem ATS0 1 D0 K0 W0 25 pin Modem Connector 1 2 3 Powe...

Page 172: ...71 72 88 89 100 111 122 123 125 128 150 external synchronization 2 40 63 71 100 129 H harmonic 1 3 4 8 10 17 56 57 58 65 66 82 100 111 112 122 123 131 134 135 137 144 I inputs 1 2 4 9 11 16 17 25 28...

Page 173: ...485 102 103 S setpoints 2 3 45 72 76 77 87 89 90 97 98 125 sliding demand 61 standards 102 103 109 status 2 3 39 40 45 70 71 72 73 77 85 86 88 92 93 96 100 107 123 136 138 T terminals ii 14 15 16 19 2...

Reviews: