Installing a Cable with Cable Gland in the Filling
Module
§
To open the filling module, remove the four screws
from the cover and remove the cover.
Note:
Make sure the seal rings are in place before you re-
close the filling module housing. Replace any defec-
tive seal rings.
§
Remove the protective cap from the bore hole on
the indicator.
Note:
The seal and nut are required for installing the
cable.
§
Guide the cable with the pre-installed cable gland
through the bore hole.
Note:
Place the seal from the protective cap between the
cable gland and the housing.
§
Install cable gland on the housing
in accordance with industry standards.
§
Please use extreme caution when performing work
on the cable gland.
Make sure IP protection is intact and the cable con-
nection withstands strain.
Attach the nut from the protective cap on the inside
of the housing. Use a torque wrench and tighten the
cable gland to 3 Nm.